26 / 54
Alpha
6.4
Connection of external circuits
Operation as circulation thermostat
Pump connection is standard for cooling thermostats and available
as accessory for immersion and heating thermostats.
Assembly for immersion/heating thermostats only:
Pull out the mains plug!
Fix the external circulation set at the bottom of the control
head by means of the two screws (pan head with cross re-
cess). The external circulation set belongs to the right side of
the head.
Hang the thermostat into the bath and fix the clamp with the
knurled thumb
screw on the edge of the bath.
When used as circulation thermostat, care for shortest hose
connections with largest inner diameter as possible. This gives
the best flow.
Connect 9 mm internal diameter tubing to the pump con-
nector.
Pump connections
– return to bath
– pump outflow or pump pressure side.
Pump connection link (silicone rubber)
If the cross-section of the tubing is too small
temperature drop between bath and external system
due to low flow rate. Increase the bath temperature appropriately.
Always ensure the maximum possible flow cross-section in the external circuit!
If the external consumer is placed at a higher level than the thermostat, the pump is stopped
and air penetrates into the thermostating circuit, the external liquid may drain down into the
bath even with a closed system
danger of flooding the thermostat!
Protect tubing with hose clips against slipping off!!
When no external consumer is connected to the thermostat, the pump outflow connection
must be linked to the return (pump connection link).
Summary of Contents for Alpha
Page 2: ......
Page 4: ......
Page 14: ...14 54 Alpha 09 2020 j 1 Control head with clamp 2 Bath vessel 3 Four mounting feet 1 2 3 ...
Page 16: ...16 54 Alpha 09 2020 j 1 Mains cable 2 Rating label 3 Cooler cable 4 Bath drain nozzle 1 2 3 4 ...
Page 47: ...09 2020 j Alpha 47 54 Cooling characteristics Alpha RA 8 RA 12 and RA 24 230 V 230 V ...
Page 51: ...09 2020 j Alpha 51 54 intentionally left blank ...
Page 52: ...52 54 Alpha 09 2020 j intentionally left blank ...
Page 54: ...54 54 Alpha 08 2020 i ...
Page 55: ......