29 / 54
7.4
Temperature setpoint setting
SEt
The setpoint is the temperature, which the thermostat should reach and maintain constant.
Press key to show
SEt
(Setpoint).
Press, display flashes.
or
Decrement or increment the Temperature Setting (
(Temperature can be set from -25 °C up to 100 °C depending on
limits “Hi” and “Lo”).
Wait 4 seconds or
Display flashes 4 seconds
new value is automatically accepted, or
value is accepted immediately with <Enter> key.
When the setpoint temperature is set below the bath temperature, it may take
up to one minute before the blue LED
lights.
Entering value closer than 5 °C to the maximum or minimum temperature limit
(Hi and Lo) are not accepted. A buzzer signal appears. If necessary, first
change the temperature limits (
7.5
Automatic Shutdown Timer
ASd
The automatic Shutdown Timer can be activated/deactivated, read out or be set. At shutdown time, the pump, heater
and compressor are deactivated, the display then shows “
StbY
” (
1x
to
ASd
Call the Automatic Shut-Down Timer
ASd
.
Read Timer:
The display shows… “
00
.
00
”:
timer not active;
The display shows… “
hh.mm
” (any value):
timer active,
(
hh.mm
minutes remaining to shutdown)
or
Set Timer:
Increment or decrement “
hh.mm
” (maximum 99:59)
Within 4 seconds after the last change of values confirm the timer value by press-
ing <Enter>.
You then proceed to Main Menu “Bath Temperature”: A blinking point indicates
that the timer is running!!
No change is done without confirmation by <Enter>!
You can deactivate the timer any time by setting the value “
00
.
00
”.
Summary of Contents for Alpha
Page 2: ......
Page 4: ......
Page 14: ...14 54 Alpha 09 2020 j 1 Control head with clamp 2 Bath vessel 3 Four mounting feet 1 2 3 ...
Page 16: ...16 54 Alpha 09 2020 j 1 Mains cable 2 Rating label 3 Cooler cable 4 Bath drain nozzle 1 2 3 4 ...
Page 47: ...09 2020 j Alpha 47 54 Cooling characteristics Alpha RA 8 RA 12 and RA 24 230 V 230 V ...
Page 51: ...09 2020 j Alpha 51 54 intentionally left blank ...
Page 52: ...52 54 Alpha 09 2020 j intentionally left blank ...
Page 54: ...54 54 Alpha 08 2020 i ...
Page 55: ......