30 / 54
Alpha
7.6
Defining temperature limits
Hi
and
Lo
With this function, it is possible to define a minimum and a maximum temperature in which the thermostat controls. By
reaching the temperature limits, a warning appears. In this way setpoint input can be prevented which may damage the
heat transfer medium or the apparatus. For example, if water is used as the heat transfer liquid, 95 °C would be practi-
cable as the maximum temperature and 5 °C as the minimum temperature. The default values are Hi = 105 °C and
Low = -30 °C.
2x for
Hi
3x for
Lo
and 1x
Select option
Hi
(maximum temperature),
or option
Lo
(minimum temperature) using the key
.
Enter the selection
The maximum temperature can be set from 35 °C to 105 °C.
The minimum temperature can be set from -30 °C to 30 °C.
The current temperature limit value is displayed flashing.
or
Set the required temperature limit.
Wait 4 seconds or
Display flashes 4 seconds
new value is automatically accepted, or
By pressing the <Enter> key the value is accepted immediately.
7.7
Refrigeration unit submenu
CooL
Menu
Cool
only with cooling thermostats.
4x
to
CooL
and 1x
Call the option
CooL
.
or
Configure the refrigeration system with these possible settings:
Auto
automatic refrigeration mode;
on
refrigeration on duty;
oFF
refrigeration out of duty.
The selected mode is displayed flashing.
Wait 4 seconds or
Display flashes 4 s
new selection is automatically accepted,
or
it is immediately accepted with the <Enter> key.
H i
L o
Summary of Contents for Alpha
Page 2: ......
Page 4: ......
Page 14: ...14 54 Alpha 09 2020 j 1 Control head with clamp 2 Bath vessel 3 Four mounting feet 1 2 3 ...
Page 16: ...16 54 Alpha 09 2020 j 1 Mains cable 2 Rating label 3 Cooler cable 4 Bath drain nozzle 1 2 3 4 ...
Page 47: ...09 2020 j Alpha 47 54 Cooling characteristics Alpha RA 8 RA 12 and RA 24 230 V 230 V ...
Page 51: ...09 2020 j Alpha 51 54 intentionally left blank ...
Page 52: ...52 54 Alpha 09 2020 j intentionally left blank ...
Page 54: ...54 54 Alpha 08 2020 i ...
Page 55: ......