PLMNL0243 REV. A, Effective Date: 03/24/16
52
FiberMINI ™ 2.0 Operation Manual
TB2 CONTINUED
PIN
FUNCTION
DESCRIPTION
10
READY
Digital Output
When active, this output indicates that the system is operating
normally. The absence of this signal indicates a problem with the
system and should be tied in conjunction with the motion inhibit
function of the controlled device. The Ready output signal will
turn off when the ground voltage is above the set limit. However,
the Ready indicator light will remain on as long as the Ready
status from the HSU is active. (<100mA)
11
TOUCH
Digital Output
When active, this output indicates that the sensor tip has come
into contact with an object. This may be the target material,
process debris or an obstruction. (<100mA)
12
STATUS
Digital Output
The Status output is a handshake for the calibration input (CAL).
When the sensor is requested to calibrate, this signal will become
active. Upon completion of the calibration, this signal will
deactivate. If the sense circuit to the tip is shorted, Status will not
become active after the request for calibration. If the sense circuit
is open, Status will remain active for 1-2 seconds after the
request for calibration. (<100mA)
LED STATUS INDICATORS
INDICATOR
DESCRIPTION
1
POWER
Power is being applied to the Basic+ Interface.
2
CAL
The CAL input is active.
3
READY
The READY signal from the HSU is active.*
4
TOUCH
The TOUCH signal from the HSU is active.
5
STATUS
The STATUS signal from the HSU is active.
6
HSU ON
The Basic+ Interface has turned power on to the HSU. Power to
the HSU is turned off if the HSU is disconnected.
7
GND OK
The ground voltage is at an appropriate level.* The Ready output
signal will be turned off when the ground voltage is above the set
limit.
*The READY and GND OK indicators will flash when the HSU is disconnected from the Basic+ Interface.