PRM21036 Reference Manual
Overview
B-1
Appendix B
Legacy Operation
Automatic Calibration Checks in SLM Utility-G3 Software ........................................B-12
B.1 Overview
This module provides information for specifying PRM2103-FF heater
controls, viewing data, performing calibration checks, and upgrading
firmware through the Model 831 and SLM Utility-G3. The Model 831 can
also be used with the latest software, G4 LD Utility.
B.2 Performing the Acoustic Calibration
To perform an acoustic calibration, follow these steps:
Step 1
Remove the birdspike and windscreen.
Step 2
Loosen the thumb screw and lift the assembly from the
PRM2103-FF to access the microphone. Do not remove the
microphone or the microphone grid cap.
Step 3
Place the calibrator over the microphone. Apply the
calibrator slowly to avoid applying a sudden large pressure
change to the diaphragm.
Step 4
To activate the Calibration function on the sound level
meter, press the
3
(TOOLS) key and highlight the Calibrate
icon.