PRM2103 Reference Manual
Performing Calibration Checks
B-9
TAKE NOTE
The standard check
level is set by performing a
manual Cal. Check directly
after performing an acoustic
calibration.
FIGURE B-9
Cal. Check Tab
The Standard Cal. Level field holds the last saved cal check value. The
default value is 94.00 dB.
You can manually change the dB level by navigating to the field and
changing the number value.
After performing the calibration check, the Model 831 displays a message
prompting you to save a new Standard Cal. Level, if needed, as shown in
Figure B-10.
TAKE NOTE
Answer Yes to save
the Standard Cal. Level value.
This should be done
immediately following the first
calibration check. Otherwise,
answer No to simply record the
level (and spectrum) in the
Calibration Check History for
trend determination.
FIGURE B-10
Save New Standard Cal. Level
B.4.2 Automatic Calibration Checks on the Model 831
Calibration checks can also be scheduled to occur automatically on the
Control tab by following these steps:
LEARN MORE
For more
information on setting up
automatic calibration checks,
refer to the Model 831 Sound
Level Meter Manual.
Step 1
Set the run mode to
Continuous
.
Step 1
Select and enable the Cal-Check option.
Step 1
Set the time of day when the check is to be performed. The
default time is 2:30 AM.
Summary of Contents for PRM2103
Page 1: ...Model PRM2103 Outdoor Microphone Preamplifier ...
Page 2: ...I2103 01 Rev C Larson Davis PRM2103 Preamplifier Reference Manual ...
Page 34: ...AEC206 Reference Manual Electrical A 4 FIGURE A 2 1 3 Octave Self Generated Noise ...
Page 91: ...PRM2103 Reference Manual Upgrading Firmware with SLM Utility G3 B 16 ...