PRM2103 Reference Manual
Performing the Acoustic Calibration
3-2
Step 3
To activate the Calibration function on the sound level
meter, navigate
3
(TOOLS)
Calibrate
.
FIGURE 3-1
Calibration
Step 4
Select a calibrator from the list or enter new information
about a calibrator.
Step 5
Turn on the calibrator. Select
Do Calibration
.
FIGURE 3-2
Calibrating Message
Step 6
When the calibration is completed, another message box
appears that shows the amount of change being made to
the calibration. To accept the calibration, select
Yes
. To
reject this calibration, select
No.
Step 7
After performing this acoustical calibration, perform a
calibration check, see “Performing Calibration Checks” on
page 3-10.