HVM200 Reference Manual
Connecting the Accelerometer
2-9
FIGURE 2-11
Sensor Settings
Step 7.
On the
Sensor List
page, do the following:
a.
Enter the model and serial number of your accelerometer.
b.
Enter the sensitivity for the x, y, and z axes.
TAKE NOTE
Sensor information, including model, serial number, and
sensitivity specifications are usually listed on the calibration certificate that
comes with an accelerometer.
c.
Click
Add
.
Step 8.
Select the accelerometer when it appears in the list and click
Select
. The
sensitivity values automatically appear on the Sensor tab.
Step 9.
Click
Save
. For future measurements, access the Sensor List page to
quickly select your configured sensors.
2.7.1 Transducer Electronic Data Sheet (TEDS)
If you are using an accelerometer with Transducer Electronic Data
Sheet (TEDS) capabilities, the x, y, and z values for sensitivity will
automatically display. Additionally, TEDS sensitivity overrides any
sensitivity values you enter manually.
If TEDS is successfully found in the accelerometer, the T icon appears in
the top display as shown below.
FIGURE 2-1
TEDS
Orange TEDS icon indicates that not all
sensitivity values are discovered.
Green TEDS icon indicates all sensitivity values
are discovered and functioning.
Summary of Contents for HVM200
Page 1: ...Larson Davis HVM200 Reference Manual...
Page 43: ...HVM200 Reference Manual A 8...
Page 44: ...HVM200 Reference Manual A 9...
Page 45: ...HVM200 Reference Manual A 10...
Page 54: ...HVM200 Reference Manual A 19...
Page 55: ...HVM200 Reference Manual A 20...
Page 56: ...HVM200 Reference Manual A 21...
Page 57: ...HVM200 Reference Manual A 22...
Page 58: ...HVM200 Reference Manual A 23...
Page 76: ...HVM200 Reference Manual E 1 Appendix E Declaration of Conformity...