HVM200 Reference Manual
A-5
FIGURE A-2
Circuit for electrical testing
Circuit to Inject Electrical Signal into HVM200 ICP
®1
Inputs
Input
• Input type: ICP, IEPE, or CCP
• Excitation current: 2 mA
• Input connector: 1/4-28 4-pin male (the
input connection is also the transducer
connection)
• Measurement input voltage range: 1.8 to 16
Vdc
• Measurement AC reference bias voltage: 9
Vdc
• Absolute voltage range (min to max): 0 to
28 V
• Bandwidth: 0.4 Hz to 3000 Hz
• Range: Single range
• Sample rate: 7161.45833 Hz
FIGURE A-3
Pinout
Represents one
input for each axis
(x,y, z).
Refer to the
“Pinout” diagram
below for more
information on
the 1/4-28 input
connection.
1. ICP is a registered trademark of PCB Piezotronics,
Inc.
Pin
Signal
A1
GND
A2
X Axis
B1
Y Axis
B2
Z Axis
A1
A2
B2
B1