96
Operating Manual
General Operating Instructions
OMBG3000 – Iss.01
© Copyright 2016
Screen 14 - Configure Alarms
2)
Press the right soft key to ‘Exit’ without saving any changes.
3)
To ‘Edit’ a parameter, press the middle soft key to enter edit mode. Using
the scroll keys, select a parameter to edit (highlighted in the table) and
select the field with the
key.
Definitions
Term
Definition
SP
The sample point (1, 2, 3, 4, or All) for which the alarm
condition is to be monitored.
>
Alarm to trigger above the value.
<
Alarm to trigger below the value.
Disabled
Alarm is disabled and will not activate.
Normal
When an alarm has occurred and the gas concentration
reaches its recovery value, the alarm will deactivate.
Latched
When an alarm has occurred and the gas concentration
reaches its recovery value, the alarm will remain activate until
cleared by the operator.
Value
The gas concentration for which the alarm condition will
become active.
Recovery
The gas concentration for which the alarm condition will
recover.
4)
Use the scroll keys to select the chosen parameter and select using the
key, or use the keypad to enter the gas concentration for the value
followed by the
key to confirm.