CHAPTER 6
PROBLEMS
NU_810-815_LX_ANG_1306
65
3.
FAULT ALARMS
These alarms stop the operation of the washer. To clear the alarm mode, the washer must be
switched off by pressing
button.
3.1 FAULT N°1 PROBE FAULT.
F
A
U
L T
N °
0 1
P
R
O
B E
D E F E C T
Probe Fault Definition :
The probe used to check the temperature of the water in the chamber is disconnected or
defective.
Corrective Action :
CONTACT A MAINTENANCE TECHNICIAN to check the electrical connection and the
condition of the probe.
3.2 FAULT N°2 EMPTYING FAULT.
F
A
U
L T
n
°
0 2
E M P T Y
I N G
F A U L T
Emptying Fault Definition :
The washer has not been fully emptied at the end of the normal emptying time defined in
the Common Times menu.
Corrective Action :
Check that there is no restriction on the emptying line, that the emptying pump (or the
valve) and its pipes are not blocked and that the chamber filters are not clogged.
This message may be displayed at the end of
PREWASH
1,
PREWASH
2,
WASH
,
RUNNING
WATER RINSE
1,
ACID RINSE
,
RUNNING WATER RINSE
2,
COLD DEMI
.
WATER RINSE
and
HOT DEMI
.
WATER RINSE
.