background image

11 

 

Manual  Mode enables to choose your measurement  parameters like «Measuring System », 
« Speed or shear rate» then « Time of measurement ». 

 

 

 

 

 

 

 

 

 

 

 

If your measuring system is not in list, you may have to create it. Please refer to section 2.2.5.6. 

Choice  between  “Speed”  or  “Shear  rate”  is  according  to  your  measuring  system.  For  MS 
ASTM, MS BV, MS KREBS and MS VANE, you should have only possibility to set the speed. 
For all other measuring system, you have to use shear rate. If you need to know what is the 
corresponding  speed  then  you  are  using  shear  rate,  you  have  to  use  constant  K

D

  of  your 

measuring system (information available in section 2.2.5.6).  

 

With speed unit in rpm, shear rate in s

-1

 and K

D

 is rpm/s

-1

Then  your  settings  are  ok,  you  must  do  a  zero.  All  LR  models  require  zeroing  before 
measurement. The zero is done with the mobile and same speed as for measure. Once the 
zero has been completed, the instrument will prompt you to place the mobile in the product for 
measurement (see section 3). 

 

 

 

 

 

 

 

Rq

 :  If   « Time » =  0,  you  could modify  « speed »  during  the  measurement. This 

could help you to define the best conditions to work on your sample. 

 

SPEED = SHEAR RATE / K

D

 

 

Select your measuring 
system 

Choose speed or shear rate 

Choose time for measurement 
(minimum 10 s) 

Summary of Contents for FIRST PLUS LR

Page 1: ...1 USING MANUAL FIRST PLUS LR...

Page 2: ...EW RESULT 13 2 2 4 REMOTE CONTROL 14 2 2 5 PARAMETERS 15 2 2 5 1 LANGUAGES 15 2 2 5 2 DATE HOUR 16 2 2 5 3 SOUNDS STANDBY LIGHTING 16 2 2 5 4 USER NAME 17 2 2 5 5 UNITS 19 2 2 5 6 MEASURING SYSTEM 19...

Page 3: ...the shear rate Those samples are named Newtonian fluids i e Oils Water Glycerol etc However many substances have a variation of viscosity in function of speed of shearing and the Flow Behaviour of tho...

Page 4: ...order this foam can be inside a box or a carrying case and can contain measuring system In detail you will find different part in your box as shown below Hardened Aluminium Stand Aluminium Arm Screw...

Page 5: ...d a clearer view of your data and analysis results On Off Switch Always with the aim of improving your experience LAMY RHEOLOGY has decided to equip all of its PLUS range with a luminous and design sw...

Page 6: ...ped with a ring with a clamping button that is used as a stop for a repeatable positioning during the measurement The rod is equipped with a white Delrin ring acting as a stop thus avoiding the arm be...

Page 7: ...ussian Spanish Compatible measuring system MS ASTM MS ULV Compatible temperature control EVA LR EVA DIN CT DIN RT 1 PLUS Supply voltage 90 240 VAC 50 60 Hz Options Carry case PN 100500 Rack stand PN P...

Page 8: ...you must first screw the rod to the base with screw and key provided Then make sure to insert the white Delrin cylinder and set it to the lowest position Then place the stop ring on the stainless ste...

Page 9: ...button to switch on your device see section 1 2 2 1 STATE ICONS Once your device is switched on you will see some icons on Touch Screen No Device is connected to the instrument Only one Device is conn...

Page 10: ...Acces is always available by clicking 2 2 2 Measure Measure tab is central part of your FIRST PLUS LR Before to use it you should install your measuring system and your sample Please see section 3 Th...

Page 11: ...re using shear rate you have to use constant KD of your measuring system information available in section 2 2 5 6 With speed unit in rpm shear rate in s 1 and KD is rpm s 1 Then your settings are ok y...

Page 12: ...to the upper limit If you change speed or spindle you will need to make a new zero adjustment Picture below show how should be the closer position regarding lower limit Then your measurement is finish...

Page 13: ...ich one you want to read Then you selected a measure you will see result and get possibility to print it if a printer is connected By clicking on Global Export you will have possibility to transfer al...

Page 14: ...By click on tab you could delete saved measure one by one as you want from your FIRST PLUS LR memory Then you click on Delete saved data will be completely deleted from internal memory without any ne...

Page 15: ...our device It is reachable by clicking on icon wheel in upper left corner of touch screen This icon is only available then you are in Main menu 2 2 5 1 Languages Enable you to select language of your...

Page 16: ...ode of your FIRST PLUS LR Choose if you want to get sound during using touch screen Choose if you want to switch off automatically your device after no using After selecting Standby enable you will ha...

Page 17: ...st account which will become the administrator and thus create or delete another operator account After specifying the name and password the administrator will be named in red in the list You can now...

Page 18: ...rning to the Main Menu you will see the name of the operator in use By clicking on the arrow below the name of the operator you can switch off the FIRST or change operator If the instrument is switche...

Page 19: ...e and corresponding constant You are not allow to change constant of existing measuring system If you want to use new constant you have to create a new measuring system with name as example Copy of an...

Page 20: ...meters and settings of your device It should be set by an administrator or responsible of the device This function is not comparable to the User name menu please see section 2 2 5 4 It should be use i...

Page 21: ...ted you will see this icone on FIRST Screen see picture below Protected mode protect programs measuring parameters and some menu as shown on pictures below 2 2 5 8 Miscellaneous This menu allow you to...

Page 22: ...he print interval time you want during measurement The FIRST can be connected to all printers with a PCL5 print protocol This includes many A4 printers The connection is made to the USB host port on t...

Page 23: ...on your viscometer Indeed some measuring systems require the installation of accessory before the insertion of the spindle As the FIRST PLUS LR get only one kind of bayonet coupling system way to inst...

Page 24: ...ity fluids and R mobiles for medium to high viscosities see tables below Designation spindle Part Number Spindle Part Number Complete setb Viscosity range for version LR mPa s RV1 111001a 111948 Not U...

Page 25: ...rpm For the measurement of a high viscosity product choose ASTM R7 and run the machine at low speed for example 1 rpm Place the measuring head in the highest position use screw on aluminium arm as wro...

Page 26: ...een using a gage Ensure that the measured torque is always sufficiently far from the lower and upper limits at least 5 above and below If this is not the case you can either change the measuring spind...

Page 27: ...trol EVA DIN PLUS and CT DIN or alone without any other accessory The part number 116031 is dedicated for measurement with disposable cup and include item 116016 111934 114436 and 114306 100 disposabl...

Page 28: ...the tube is properly installed check that the temperature sensor on the viscometer is at least one centimetre deep into the liquid When using your viscometer with a temperature control system EVA MS...

Page 29: ...ng careful if the temperature is high You can then remove the cylinder from the tube to clean it Remove the cap from the tube to clean it Clean the temperature sensor 3 3 2 Use of Item 116031 The meas...

Page 30: ...r When the measuring head is properly positioned there is no need to lock the position using the screw on the arm Also be sure to allow sufficient time for stabilization if the measurement temperature...

Page 31: ...nstall spindle L2 See section 2 2 2 Select spindle L2 and speed 60 rpm Make a zero adjustment See section 3 2 and immerge the spindle in the oil at the good level mark on the spindle Wait for 15 minut...

Page 32: ...32 LAMY RHEOLOGY 11 A rue des Aulnes 69410 Champagne au Mont d Or France Tel 33 0 4 78 08 54 06 Fax 33 0 4 78 08 69 44 contact lamyrheology com...

Reviews: