22
2.2.5.9.
Density
Enable you to enter density value of your product in order to measure his kinematic viscosity.
If you set a density value, you will get all the time viscosity in cStoke. Please remove density
information if you want to get back Pa.s or Poise for unit of viscosity.
2.2.5.10. Printing
Allows you to connect a printer, print a test page, and choose the print interval time you want
during measurement.
The FIRST can be connected to all printers with a PCL5 print protocol. This includes many A4
printers. The connection is made to the "USB host" port on the back of the instrument.
Once the printer is connected, simply click on "Install Printer".
Then you print data at the end of measure or a saved file, you will have on only information
shown on device screen as final result. If you want to have more data printed, you have to
select
“Interval printing” time to get data printed between start and end of your measurement.