3

3.  Do not under fill the disk. If the disk is under filled, 

the blank chamber will not be full. The reagent 

chambers will not fill entirely and results will be 

inaccurate. 

Underfilled 

Blank 

Chamber

4.  Do not introduce air bubbles into the disk. The 

reagent chambers will not fill entirely and results 

will be inaccurate. As soon as a bubble starts to 

form, pull back on the plunger to draw the bubble 

out of the disk. Begin the filling process again.

Air Bubble

5.  Wet disks should be dried thoroughly with a lint 

free wipe. Handle the disk by the edges. Between 

tests store the disk on a flat surface. Do not 

agitate a partially filled disk to avoid leaking and 

contamination. Partially used disks should be 

discarded at the end of the day.

6.  Disks should be filled and used within 10 minutes. 

They cannot be filled ahead of time.

For filling and troubleshooting tips go to: 

lamotte.com/spin_support

GENERAL OPERATING PROCEDURES

METER

  When a filled disk is placed in the chamber and the lid is closed, the meter 

spins at high speed to distribute the sample to the test wells. Next the meter slows 

to maximize the pumping action of the stainless steel mixing beads as the reagents 

mix with the sample water. Each reaction is then read at the proper time and 

wavelength for that reagent system.

The button located in the lower center of the top of the meter turns the 

instrument on and off.

The Blue indicator light of the on/off button indicates the status of the instrument. 

Steady blue light 

 the blue LED will remain steady to indicate that the meter is on 

and ready to run a test.

Blinking blue light

 (three blinks/second) – a test is in progress and the disk is 

spinning. Do not open the lid when the disk is spinning. 
Care should be taken when closing the lid. Do not slam the lid. Wiring between the 

lid and the body of the photometer passes through the hinge. The meter will not run 

with the lid open.

TOUCHSCREEN DISPLAY

When the button is pressed to turn the meter on, the Test Screen will be displayed.
The display screen is touch-activated. To make a selection, tap the icon or word on 

the screen with a fingertip, fingernail, pencil eraser, or stylus.

 

¬

Gently wipe smudges from the screen with the Cloth Wipe (3580-WIPE).

 

¬

Do not touch the screen with a sharp object.

 

¬

Do not place objects on the screen that will scratch or damage it.

 

¬

Avoid touching the screen with wet fingers.

TESTING

1.  Press 

 and hold until 

the meter turns on.

2.  Tap  . Select a sanitizer 

type. Tap   to confirm.

3.  Tap  . Select a disk 

series (found on disk 

packaging). Tap   to 

confirm. 

NOTE: Disk Series are 

limited by Sanitizer 

selection or Water Source 

selection.

4.  Remove a disk from the 

packaging.

5.  Use the syringe (1189) to 

fill the disk with the pool 

or spa water sample.

6.  Insert the disk. Cover the 

disk with the Universal Disk Cover (1719). Close the lid.

7.  Tap 

 to start test. Tap 

 to cancel the test. If the test is cancelled discard 

the disk.

8.  The results will be displayed. 
9.  Choose an option.

 

¬

Tap 

 to add a tag.

 

¬

Tap the highlighted   to save the test results to the test log if Auto Save is 

not enabled.

 

¬

Tap the hightlighted 

 to send the results to the enabled Mobile Bluetooth 

Printer.

 

¬

Tap the highlighted 

 to send the results to a Bluetooth enabled device.

 

¬

Tap   to return to the Test Screen.

10.  REMOVE THE DISK FROM THE CHAMBER. DISPOSE OF THE DISK. Disks left in the 

chamber can leak and damage the device. Do not store the disk cover on the 

hub.

11.  Press and hold 

 for 2 seconds to turn the meter off.

NOTE: For water samples over 100 °F (38 °C) subtract 0.1 from pH result or, for the 

most accurate result, wait until water sample is below 90 °F (32 °C) to test.

SAVE, PRINT, AND SEND TEST RESULTS

 

Test results can be saved, transferred to 

the WaterLink Connect 2 mobile app via 

Bluetooth, and sent to the BLE Mobile Printer 

(Code 5-0067).

Test History Settings  

The WaterLink Spin 

Touch can log test results for 250 water 

samples in the Test History. The results for 

the most recent sample will be located at 

the top of the list. All results can be logged 

automatically or results for an individual 

sample can be logged after the sample has 

been tested. 
To turn on automatic logging, tap 

 on the 

Test Screen. Tap  . Select Auto-Save Tests. 

Tap 

 and   to return to the test screen. When Auto Save is selected   will not be 

highlighted on the Test Results screen.
To manually log results for one sample at a time, Auto-Save Tests must be disabled. 

If Auto-Save Tests is disabled   will be highlighted on the Test Results screen. After 

the test has been run, tap   to save the results for that water sample to the Test 

History. 
A user defined identification tag can be assigned to an individual sample result. 

To enable tagging, tap   on the Test Screen. Tap 

. Select Enable tags. Tap 

 

and   to return to the test screen. The tag can be up to 4 hexadecimal characters. 

It cannot begin with a zero. To tag a test result, tap 

 on the bottom of the Test 

Results screen and use the buttons to enter up to four characters. Tap 

 to erase 

a character. Tap 

 to save the tag and return to the Test Results screen. Tap 

 to return to the Test Results screen without saving the tag. The tagged result 

will appear on the Test Results screen and the Test History screen. The tag will be 

preceded by “T-“.
Logged results are viewed in Test History. Controls for viewing and managing single 

or multiple test records are located on the Test History screen. Tap the checkbox 

next to a test record to select it, then tap one of the buttons along the bottom to 

perform an action with the selected records.

Print or Send Results 

to a Bluetooth Phone 

or Tablet

 

 The WaterLink 

Spin Touch can print to 

the Mobile Bluetooth 

Printer or send test 

results to a Bluetooth-

enabled device such as 

a phone or a tablet. The 

highlighted 

 or 

 

will show which function 

is active. The Spin Touch 

cannot connect to the 

printer and connect to 

a phone or tablet at the 

same time. 
There are two modes of operation for connecting to the printer and a phone or tablet 

– Fast Printer Connect 

ON

 and Fast Printer Connect 

OFF

. The default mode is Fast 

Printer Connect 

ON

. Fast Printer Connect can be turned on or off from the Bluetooth 

Menu that is located in the Settings Menu.
If Fast Printer Connect is 

ON

, the meter will connect immediately to the printer upon 

recognizing it. It will also connect to a phone or tablet but priority is given to the 

printer connection. The printer will start printing immediately when 

 is tapped. 

The 

 will dim in intensity while the printer is printing. Fast Printer Connect should 

be 

ON

 if results will be printed more frequently than results will be sent to a phone or 

tablet. It can remain selected if results will only be sent to a phone or tablet. If Fast 

Print Connect is 

ON

, and the printer is on, the printer will have to be turned off before 

results can be sent to a phone or tablet. 
If Fast Printer Connect is

 OFF

, the meter has the opportunity to connect to a phone 

or tablet at all times other than when the printer is actually printing a test. The meter 

will connect to the printer only after 

 has been tapped so printing will be delayed a 

few seconds while the connection to the printer is being established. The 

 will dim 

in intensity while the printer is printing. After printing, the meter will automatically 

reconnect to the phone or tablet. Fast Printer Connect should be 

OFF

 if results will be 

sent to a phone or tablet most frequently and printed less often.

SYRINGE

  

A plastic 3 mL syringe (Code 1189) is used to fill the disks. A 

precision tip on the syringe fits into the fill hole on the disk. The syringe tip 

should not be removed from the syringe. Syringes should be cleaned between 

water samples. Pump air in and out of the syringe a few times to clear the 

previous sample or rinse the syringe with a small amount of the next water 

sample before filling it with the next sample. Replace syringes when the 

tips become worn or the plungers don’t move smoothly. See Accessories and 

Replacement Parts.

REAGENT DISK

 

 The WaterLink Spin Touch uses a SpinDisk™ 

reagent system. The dried reagents are packaged in single test 

amounts in a sealed, polystyrene disk. Stainless steel mixing beads 

12:14 PM

16-Apr-20

Sani

ti

zer

Disk Series

Salt

203

Tap to

Start Test

Sanitizer

Start

Test

Disk

Type

Power/

Battery

Status

Date &

Time

Test

History

Settings

Test Time 05:30:06 16-Apr-2020

Test Results

Transfer

Results

via

Bluetooth

Return

to Test

Screen

 

FCL - 1.31

TCL - 1.31

CCL - 0.00

PH - 7.8

ALK - 100

HARD - 230

CYA - 2

COPPER - 0.1

IRON - 0.1

PHOS - 0

SALT - 489

Save 

Results

Print

Results

via

Bluetooth

Tagging

Select All/

De-Select All

12:02A 16-Apr Salt 401
12:01A 16-Apr Salt 301
12:00A 16-Apr T-4650 Chlorine 100
12:00A 16-Apr Chlorine 300
11:57A 16-Apr T-ABC7 Chlorine 300
11:49A 16-Apr Biguanide 400

Test History

View 

Selected 

Results

Delete

Selected

Results

History

Settings

Select 

Auto Save

Go to Top

Go to Bottom

Scroll up 

or down

the test

list by 5  

entries

Home

 

Summary of Contents for WaterLink SpinTouch 3580-AU

Page 1: ...SF ANSI CAN 50 2023 LaMotte Company All Rights Reserved For Code 3580 AU 3581 AU 05 09 2023 Scantoaccess download andprint theSpinTouchQuickStartGuide WARNING only use the USB cable and wall adapter t...

Page 2: ...a mobile device requires that a mobile app is installed on the device and an active account for the associated software program For example with an active WaterLink Solutions account and the WaterLink...

Page 3: ...at a time Auto Save Tests must be disabled If Auto Save Tests is disabled will be highlighted on the Test Results screen After the test has been run tap to save the results for that water sample to th...

Page 4: ...d at low battery power without connecting it an AC power source the meter will go into an auto shutdown mode In this mode the meter will be locked until meter is connected to an AC source and the batt...

Page 5: ...k Free Chlorine Bromine Free Chlorine Bromine Total Chlorine Total Chlorine Chlorine Bromine pH Alkalinity Code4335 Series601 pH Free Chlorine Bromine Alkalinity pH Free Chlorine Bromine Alkalinity Bl...

Page 6: ...IMPORTANT DonotattempttoseparatethecomponentsoftheMeterCheckDisk Code1705 1705 EU TheMeterCheckDiskconsistsofadiskwithapermanently attachedcover DonotfilltheMeterCheckDiskwithwater Waterisnotusedinth...

Page 7: ...emergency line for ChemTel at 1 800 255 3924 USA Canada Puerto Rico For locations outside of the North American continent call 813 248 0585 collect Ensure that the protection provided by this equipmen...

Reviews: