SPECIFICATIONS
Instrument Type
Centrifugal Fluidics Photometer
Wavelengths
(interference filters)
390 nm, 428 nm, 470 nm, 525 nm, 568 nm, 635 nm
Display
Color Capacitive Touchscreen, 3.5 in, 320 x 240 pixel resolution
Wavelength Accuracy
±2 nm
Wavelength Bandwidth
10 typical
Photometric Range
-2 to 2 AU
Photometric Precision
±0.01 AU at 1.0 AU
Photometric Accuracy
±0.01 AU at 1.0 AU
Sample Chamber
Accepts prefilled disk
Light Source
6 LEDS
Detectors
6 silicon photodiodes
Pre-Programmed Tests
Yes, with automatic wavelength selection
Languages
English, French, Spanish, German, Dutch, Swedish, Portuguese,
Italian, Chinese, Turkish
Temperature
Operation: 0-50 °C; storage – 40-60 °C
Operation Humidity Range
0- 90 % RH, non-condensing
Communication
USB-C, Bluetooth low energy technology (BLE)
Calibration
Factory set, field calibration via internet connection
Firmware
Internet updateable (New Test, New Test Calibrations, etc.)
Software
Android & iOS Apps, WaterLink® Solutions, WaterLink® Connect 2
Power Requirements
USB wall adapter, USB computer connection or internal lithium
ion rechargeable battery
Battery Type
Lithium ion
Minimum Capacity
12 V/2.6 AH
Charge Life
Approximately 150 tests
Battery Life
Approximately 500 charges
Full Charge
10-12 hours
Water Resistance
Rubber over-molded base, rubber USB Port Plug, gasketed
display and hinge.
Electrical Rating
Rated voltage (5V), Rated power of input current (1.6 A) at USB C
Summary of Contents for WaterLink 3585