Other Settings
Units allows the selection of units for the Hardness test results to be
reported as gpg or ppm for the Drinking Water market. It also allows
the selection of the Nitrate test results to be reported as ppm Nitrate
(NO
3
) or ppm Nitrate Nitrogen (NO
3
-N). When results are reported as
ppm Nitrate (NO
3
) the nitrite results will be reported as ppm Nitrite
(NO
2
). When results are reported as nitrate nitrogen NO
3
-N) the
nitrite results will be reported as ppm Nitrite Nitrogen (NO
2
-N).
About… lists the Serial Number, Firmware Version, Bluetooth MAC
address, Bluetooth Version and Test Count. The Test Count shows
the number of complete tests that have been performed over the
lifetime of the meter. Tap
to return to the Settings menu.
Ranges Enabled allows the option of having test results that are
out of the range of the reagent system displayed in red. The default
setting is on.
When Show Market Type is selected, the market type will be displayed
on the Testing Screen. The default setting is OFF. Tap to return to
the Settings menu.
Select Disk Detection to display an error message and abort the
test when improper testing conditions, such as No Disk, No Disk
Cover, Used Disk or Meter Check Disk, are present that would result
in inaccurate or no test results. The default setting is OFF. Tap to
return to the Settings menu.
RANGES
Test Factor
Range
Display Abbreviation
Alkalinity, Total
0–250 ppm
ALK
Chlorine, Free
0.00–15.00 ppm
FCL
Chlorine, Total
0.00–15.00 ppm
TCL
Combined Chlorine
0.00–15.00 ppm
CCL
Copper
0.0-6.0 ppm
COPPER
Hardness, Total
0–70 gpg/0-1200 ppm
Hardgpg/T HARD
Iron, Ferrous, soluble
0.0–6.0 ppm
FERROUS
Iron, Ferric, soluble
0.0–6.0 ppm
FERRIC
Iron, Total, soluble
0.0-6.0 ppm
T IRON
Nitrate
0-45 ppm/0-11 ppm
NITRATE/NITRATE-N
Nitrite
0.0-2.0 ppm/0.0-0.6 ppm
NITRITE/NITRITE-N
pH (4336)
6.4-10.0
pH
pH (4337)
4.5-8.6
pH
Go to SETTINGS>Other Settings to select hardness and nitrate units.
Copper concentrations over 4 ppm will result in low total hardness results.
Test results that are out of range of the reagent system wil be
RED
.
RED
test results may
not be accurate. Go SETTINGS>Other Settings to turn the Ranges Enabled feature OFF.
Not intended for testing ultra-pure samples such as samples from reverse osmosis
systems.
GENERAL OPERA
TING PROCEDURES