Lambrecht EOLOS-IND Manual Download Page 5

Manual
Weather Sensor EOLOS-IND Modbus 

(1643) EOLOS-IND Modbus 

  

 

      Manual

pointing to the north is integrated to be set into a correspond-
ing slot of the mounting pipe (if available). Thus the sensor is 
safely attached. If needed you can turn in or unscrew the pin 
by means of allen key.

To set up the sensor’s north orientation select a land-
mark which is as far as possible up north with regard 
to the fi nal position of the wind direction sensor.

The reference point can be selected using a topographical 
map (1:25000). The exact position of the reference point is 
determined using an amplitude compass that can be adjusted 
horizontally on a stand.

Compass declination has to be considered!

To align the sensor ahead (on ships) locate a point outside 
the ship in the landscape which is located in the ship ahead 
direction respectively in the centre line or in case of the sen-
sor is mounted far away from the middle line a line parallel 
to the centre line. 
Once the sensor is adjusted, it can be fi xed with the two hex-
agonal screws. Finally the earth screw has to be connected to 
the ship’s ground. Acid-free contact grease is recommended 
to protect contact surfaces against corrosion.

Follow all safety instructions while setting up the 
sensor onto a mast.

2.7.2 

Power and Signal 

 

Connection

A 4-pin/ M12 cable socket is required for the electrical con-
nection of the sensor. The shielding of the cable has to be 
connected to the protective conductor

To reduce the risk of inductive interference the sen-
sor must be properly grounded (screening on both 
sides).

The external connection is via central connector which is 
located in housing base. For further details about electrical 
connection please see chapter „Connecting diagrams“. 

If the sensor is mounted in correct manner and connected 
with the right cable (accessory), you can attach the wires to 
power supply and signal outputs to data acquisition equip-
ment (computer). 

The typical power supply requirements of the EOLOS-IND 
sensor are 24 VDC with a maximum current drain of 2.5 A. 
The input range is max. 18... 32 VDC

The output signal of the sensor corresponds to the RS-485 

standard. The signal levels allow transmission via shielded 

signal cables up to a maximum length of 1,200 meters or 

4,000 feet. The cable lengths depend on the quality of the 

cables used.

When the power supply of the sensor is switched on, after 
30 seconds (in order to reach operating temperature) the 
sensor cyclically starts sending data protocols.

2.7.3 

Safety Regulations

Because the wind sensor often is mounted on ex-
posed locations in dangerous heights the installation 
personnel has to pay attention to the relevant safety 
regulations for such works. During the electrical 
installation and termination works the external circuit-
breaker must be switched o

 .

It is not permitted to open those housings by unau-
thorised persons!

Maintenance

3.1 

Regular Maintenance and  

 

Calibration

The sensor EOLOS-IND is service reduced and designed for 
a very long lifetime. Recommended is a regular visual check 
regarding dirt of surface caused by the weather and if so, to 
clean up.

If reference measurements should be necessary 
stringently must be noted that a comparability of the 
measured values is given only if the measurements 
take place under same conditions. I.e. the reference 
equipment must be used very close to the sensor!

The sensor is a measuring instrument and thus apply user 
specifi c standards regarding period of recalibration. 
Recommendation: 2 years.
We recommend to have the fi lter cap of the humidity-tem-
perature sensor every two years exchanged in our plant. 
Depending on the application shorter maintenance cycles 
could be necessary. 

3.2 

Visual Check and Cleaning

The use of the sensor under the respective environmental 
conditions requires certain steps. It is thus recommendable 
to clean the outside of the housing and shelter within specifi c 
intervals. The intervals are dependant on the environmental 
conditions and the degree of soiling. We recommend a regular 
sight and functional check.
In case you should be faced with any specifi c problems please 
contact the LAMBRECHT meteo service under:
Tel.: 

+49-(0)551-4958-0

Fax: 

+49-(0)551-4958-327

E-Mail:  [email protected]

Transports

In case it is necessary to ship or to transport the sensor must 
be carefully packed to prevent damages during transport.

5

Summary of Contents for EOLOS-IND

Page 1: ...egrated overvoltage protection Simple space saving assembly on 50 mm standard pipe The Perfect Weather Sensor For industrial use as well as under extreme environmen tal conditions 5 1 1 parameters in...

Page 2: ...Mounting 2 7 1 North Alignment of Wind Sensor 2 7 2 Power and Signal Connection 2 7 3 Safety Regulations 3 Maintenance 3 1 Regular Maintenance and Calibration 3 2 Visual Check and Cleaning 4 Transport...

Page 3: ...is method Lightweight mass and immovable measuring elements to enable very low starting values distance and attenuation constants as well as a very high repetition accuracy The sensor can be installed...

Page 4: ...t the mast if available If needed you can turn in or unscrew the pin by means of allen key 2 7 1 North Alignment of Wind Sensor For wind direction measurements the north mark on the sen sor must be al...

Page 5: ...signal of the sensor corresponds to the RS 485 standard The signal levels allow transmission via shielded signal cables up to a maximum length of 1 200 meters or 4 000 feet The cable lengths depend o...

Page 6: ...ngsscraube M8 fastening screw Erdscraube M6 earthing screw 10 1 5 6 0 1 P40 min P50 P48 3 5 6 0 1 10 1 min 40 P40 min 50 9 P 0 1 Aufnahmerohr f r den Sensor sensor seat pipe N Schlitz im Mastende baus...

Page 7: ...ensor EOLOS IND Modbus 1643 EOLOS IND Modbus Manual 7 IN 24 V DC f r Elektronik for electronic Schirm shield Erde earth 287 56 0RGEXV 578 1 2 3 4 1 2 3 4 br wt bl bk PIN color 5 5 N A br ws bl sw Farb...

Page 8: ...is transmitted in the order 0x12 0x34 Example big endian 32 bit or 64 bit Register size value 32 bits 0x12345678 is transmitted in the order 0x12 0x34 0x56 0x78 6 2 Device Address The addresses 1 247...

Page 9: ...stantaneous value r h 10 1 decimal place INT 30602 Humidity mean value since last retrieval r h 10 1 decimal place INT 30603 Humidity maximum value r h 10 1 decimal place INT 30604 Humidity minimum va...

Page 10: ...HT meteo o er the possibility of autocon guration This is supported e g by the data logger met LOG For autocon guration the register addresses of the measured values and sensor data available in the r...

Page 11: ...imal places LONG 0x04 Big Endian WORD 31102 Precipitation amount total Low WORD 3 decimal places Registers 31101 31102 can only be read out together function code 0x04 0x04 Big Endian WORD 31201 Preci...

Page 12: ...s 1 Dew point maximum value INT 46020 30703 Register address 1 Dew point minimum value INT 46021 30801 Register address 1 Air pressure instantaneous value INT 46022 30802 Register address 1 Air pressu...

Page 13: ...100 r h Parameters Wind direction Measuring range 0 360 Accuracy 3 RMS Resolution 1 Wind speed Measuring range 0 1 50 m s Accuracy 0 5 m s 5 RMS of the meas value at 5 1 40 m s Resolution 0 1 m s Air...