2

Model 240 Quick Start Guide

Safety Precautions

Observe these general safety precautions during all phases of 
instrument operation, service, and repair. Failure to comply with 
these precautions or with specific warnings in the user’s manual 
violates safety standards of design, manufacture, and intended 
instrument use. Lake Shore Cryotronics, Inc. assumes no liability 
for customer failure to comply with these requirements.

The 240 Series input module protects the operator and 
surrounding area from electric shock or burn, mechanical 
hazards, excessive temperature, and spread of fire from the 
instrument. Environmental conditions outside of the conditions 
below may pose a hazard to the operator and surrounding area.

 

„

Indoor use

 

„

Altitude to 2000 m

 

„

Temperature for safe operation: -20 °C to 50 °C

 

„

Maximum relative humidity: 95% non-condensing

 

„

Power supply voltage fluctuations not to exceed  
±10% of the nominal voltage

 

„

Overvoltage category II

 

„

Pollution degree 2

 

„

IP20: not protected against harmful ingress of water 

 

Ground the instrument

 

To minimize shock hazard, the instrument is equipped with a 
grounded connection to the DIN rail. Connect the DIN rail to an 
electrical ground.

Do not operate in an explosive atmosphere

 

Do not operate the instrument in the presence of flammable gases or 
fumes. Operation of any electrical instrument in such an environment 
constitutes a definite safety hazard.

Do not substitute parts or modify instrument

 

Do not install substitute parts or perform any unauthorized 
modification to the instrument. Return the instrument to an 
authorized Lake Shore Cryotronics, Inc. representative for service and 
repair to ensure that safety features are maintained.

Cleaning

 

Do not submerge instrument. Clean only with a damp cloth, 
exterior only.

Installation

 

When installing the instrument, ensure it is mounted securely on the 
DIN rail.

Improper Use

 

If the instrument is used in a manner that is not specified by Lake 
Shore, the safety protections provided by the instrument are no 
longer guaranteed, and may be impaired.

For more safety information, refer to the Model 240 user’s manual:  

https://www.lakeshore.com/240manual

.

Summary of Contents for 240

Page 1: ...1 https www lakeshore com 240 Model 240 Quick Start Guide Quick Start Guide 240 Series Cryogenic Sensor Input Module...

Page 2: ...rotected against harmful ingress of water Ground the instrument To minimize shock hazard the instrument is equipped with a grounded connection to the DIN rail Connect the DIN rail to an electrical gro...

Page 3: ...ormation for getting started with your input modules For further information please see the Model 240 user s manual Front mounted OLED screen Cryogenic temperature sensor inputs 5 pin connectors Power...

Page 4: ...slightly with shipping companies Keep all damaged shipping materials and contents until instructed to either return or discard them 4 Open the shipping container and keep the container and shipping ma...

Page 5: ...oltage and current leads should then be covered with a braided or foil shield that is connected to the shield pin of the instrument This type of cable is available through local electronics suppliers...

Page 6: ...e USB disconnect it from the computer before beginning PLC communication Physical Installation The 240 Series input module has been specifically designed to mount on a DIN rail To minimize shock hazar...

Page 7: ...his is just the beginning of what is possible with the Model 240 For additional features please see the Model 240 user s manual Basic operation 1 Connect a power adapter to the input module 2 Connect...

Page 8: ...ck Start Guide 6 Turn Sensor Excitation ON Turn Sensor Excitation off to disable any unused inputs 7 Load the sensor calibration file under Sensor setup that matches the sensor by navigating to the ap...

Page 9: ...acy or High speed NOTE The Current Reversal option is available if a resistive sensor is connected and the High accuracy mode is selected NOTE High speed ID is only available for the Model 240 2P 10 U...

Page 10: ...exactly with the markings on the dial Please do not use the sensor simulator as a method for verifying accuracy of the 240 Series input module The evaluation kit includes two unique accessories for t...

Page 11: ...the terminal connector when connecting and disconnecting to the 240 Series input module 3 Plug the terminal connector into the desired input of the 240 Series input module The switch should face towar...

Page 12: ...ort The Lake Shore Forum is also a great place to look for solutions to post issues and to share successes https forums lakeshore com For further documentation and information please see https www lak...

Reviews: