9
https://www.lakeshore.com/240/
Model 240 Quick Start Guide
If you have a custom calibration curve to load,
choose Custom Curve and load your file (the Model
240 supports the .340 file format).
If a sensor with a standard curve is connected,
choose Standard Curve and select the curve from
the drop down menu.
If you aren’t using a curve, the module will continue
to monitor sensor units.
Type of sensor
Custom curve
Cernox®
Standard curve
Platinum/Diode
No curve
Simulated
8. Once the curve is loaded, rename the Input Label (if
desired).
9. Under Operating mode, select High accuracy or High
speed.
NOTE:
The Current Reversal option is available if a
resistive sensor is connected, and the High accuracy
mode is selected.
NOTE:
High speed ID is only available for the Model
240-2P.
10. Under Reporting units, select either Sensor units or a
temperature unit.
11. Live readings from your connected sensors should
now be displayed at the right of the screen, as well
as on the front OLED display of the 240 Series input
module.
12. Click
Module
to go to the Module Settings screen,
and set the PROFIBUS address.
NOTE:
For further information on communication
settings, see section 4.3.3.2 of the Model 240 user’s
manual.
13. Your module should now be configured to respond
to PROFIBUS requests, and you may disconnect the
Model 240 by clicking the
Disconnect
button.