6

Model 240 Quick Start Guide

NOTE: 

Make sure to properly seat connectors.  

When removing a sensor, use a screwdriver to gently 
remove the connector.

USB Connection

MeasureLINK™ utility

Lake Shore provides a utility called MeasureLINK™, 
which makes loading temperature curves into the 
Model 240 a very simple process. The program 
configures the module and copies curves from 
formatted files into the Model 240.

Configure the USB/PC connection

The Model 240 is fully-configurable through direct USB 
connection.  

NOTE:

 After configuring the USB, disconnect it from the 

computer before beginning PLC communication.

 

Physical Installation

The 240 Series input module has been specifically 
designed to mount on a DIN rail. To minimize shock 
hazard, the instrument is equipped with a grounded 
connection to the DIN rail. Mount the input module 
vertically, and connect the DIN rail to an electrical 
ground. If the input module is mounted in an equipment 
rack or enclosure, ensure that the ambient temperature 
does not exceed the instrument rating. 

NOTE: 

For best noise and accuracy performance, install the 

input module in a grounded metallic enclosure.

Power and RS-485

  

connections may be  
shared between  
modules by using  
the optional  
backplane  
connectors supplied  
with the module.

Summary of Contents for 240

Page 1: ...1 https www lakeshore com 240 Model 240 Quick Start Guide Quick Start Guide 240 Series Cryogenic Sensor Input Module...

Page 2: ...rotected against harmful ingress of water Ground the instrument To minimize shock hazard the instrument is equipped with a grounded connection to the DIN rail Connect the DIN rail to an electrical gro...

Page 3: ...ormation for getting started with your input modules For further information please see the Model 240 user s manual Front mounted OLED screen Cryogenic temperature sensor inputs 5 pin connectors Power...

Page 4: ...slightly with shipping companies Keep all damaged shipping materials and contents until instructed to either return or discard them 4 Open the shipping container and keep the container and shipping ma...

Page 5: ...oltage and current leads should then be covered with a braided or foil shield that is connected to the shield pin of the instrument This type of cable is available through local electronics suppliers...

Page 6: ...e USB disconnect it from the computer before beginning PLC communication Physical Installation The 240 Series input module has been specifically designed to mount on a DIN rail To minimize shock hazar...

Page 7: ...his is just the beginning of what is possible with the Model 240 For additional features please see the Model 240 user s manual Basic operation 1 Connect a power adapter to the input module 2 Connect...

Page 8: ...ck Start Guide 6 Turn Sensor Excitation ON Turn Sensor Excitation off to disable any unused inputs 7 Load the sensor calibration file under Sensor setup that matches the sensor by navigating to the ap...

Page 9: ...acy or High speed NOTE The Current Reversal option is available if a resistive sensor is connected and the High accuracy mode is selected NOTE High speed ID is only available for the Model 240 2P 10 U...

Page 10: ...exactly with the markings on the dial Please do not use the sensor simulator as a method for verifying accuracy of the 240 Series input module The evaluation kit includes two unique accessories for t...

Page 11: ...the terminal connector when connecting and disconnecting to the 240 Series input module 3 Plug the terminal connector into the desired input of the 240 Series input module The switch should face towar...

Page 12: ...ort The Lake Shore Forum is also a great place to look for solutions to post issues and to share successes https forums lakeshore com For further documentation and information please see https www lak...

Reviews: