Model 240 Quick Start Guide

Fixed resistor board

A small PCB containing two different fixed resistors 
that correspond with an arbitrary Cernox® sensor at 
approximately 1.8 K and a 100 

W

 platinum sensor at 

approximately 65 K. The evaluation kit contains enough 
of these boards for both the 2 and 8 input modules 
included in the kit so that valid readings can be easily 
obtained on all inputs. Follow these steps to use the fixed 
resistor board:
1.  Insert one of the fixed resistor boards into one of the 

5-pin sensor input connectors provided with the 240 
Series input module. Locate the board so that the 
switch is on the same side as the screw heads on the 
terminal connector.

2.  Screw down the terminal connections so that the 

terminals contact the metal traces on the fingers of 
the fixed resistor board. 

NOTE: 

The fingers can be fragile. Please handle 

the assembled set by the terminal connector when 
connecting and disconnecting to the 240 Series input 
module.

3.  Plug the terminal connector into the desired input of 

the 240 Series input module. The switch should face 
toward the front of the module.

4.  In MeasureLINK™, choose the correct input.
5.  Under Sensor Setup, select the sensor you wish to 

simulate from the drop down list under Standard 
Curve:

 

„

For Cernox® reading 1.8K, choose Sensor 
Simulator – NTC

 

„

For Platinum reading 65 K, choose LSCI PT-100

6.  Make any other changes to the input configuration 

needed for your application, and then disconnect 
from MeasureLINK™.

7.  Set the switch on the fixed resistor board to 

CERNOX if you loaded the Sensor Simulator – NTC 
temperature curve. Set the switch to PLATINUM if 
you selected the LSCI PT-100 curve instead. 

NOTE: 

If the switch is set to PLATINUM when the Sensor 

Simulator – NTC curve is loaded, the temperature 
displayed will be approximately 400 K. If the switch is 
set to CERNOX when the LSCI PT-100 curve is loaded, 
the input module will report an S.Over error since the 
resistance value is greater than the 1 kW limit for a PTC 
sensor.

Summary of Contents for 240

Page 1: ...1 https www lakeshore com 240 Model 240 Quick Start Guide Quick Start Guide 240 Series Cryogenic Sensor Input Module...

Page 2: ...rotected against harmful ingress of water Ground the instrument To minimize shock hazard the instrument is equipped with a grounded connection to the DIN rail Connect the DIN rail to an electrical gro...

Page 3: ...ormation for getting started with your input modules For further information please see the Model 240 user s manual Front mounted OLED screen Cryogenic temperature sensor inputs 5 pin connectors Power...

Page 4: ...slightly with shipping companies Keep all damaged shipping materials and contents until instructed to either return or discard them 4 Open the shipping container and keep the container and shipping ma...

Page 5: ...oltage and current leads should then be covered with a braided or foil shield that is connected to the shield pin of the instrument This type of cable is available through local electronics suppliers...

Page 6: ...e USB disconnect it from the computer before beginning PLC communication Physical Installation The 240 Series input module has been specifically designed to mount on a DIN rail To minimize shock hazar...

Page 7: ...his is just the beginning of what is possible with the Model 240 For additional features please see the Model 240 user s manual Basic operation 1 Connect a power adapter to the input module 2 Connect...

Page 8: ...ck Start Guide 6 Turn Sensor Excitation ON Turn Sensor Excitation off to disable any unused inputs 7 Load the sensor calibration file under Sensor setup that matches the sensor by navigating to the ap...

Page 9: ...acy or High speed NOTE The Current Reversal option is available if a resistive sensor is connected and the High accuracy mode is selected NOTE High speed ID is only available for the Model 240 2P 10 U...

Page 10: ...exactly with the markings on the dial Please do not use the sensor simulator as a method for verifying accuracy of the 240 Series input module The evaluation kit includes two unique accessories for t...

Page 11: ...the terminal connector when connecting and disconnecting to the 240 Series input module 3 Plug the terminal connector into the desired input of the 240 Series input module The switch should face towar...

Page 12: ...ort The Lake Shore Forum is also a great place to look for solutions to post issues and to share successes https forums lakeshore com For further documentation and information please see https www lak...

Reviews: