Technical Description
37
Electromagnetic Emissions Tests
Emissions Test
Standard or test method
Compliance
RF emissions
CISPR 11
Group 1 Class B
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/ flicker
emissions
IEC 61000-3-3
Complies
Electromagnetic Immunity Tests
Immunity Test
Standard or test method
Compliance Level and
Immunity Test Level
Electrostatic discharge
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV,
± 15 kV air
Radiated RF EM fields
IEC 61000-4-3
3 V/m
80 MHz – 2.7 GHz
80% AM at 2 Hz
Proximity fields from RF
wireless communications
equipment
IEC 61000-4-3
380-390 MHz: 27 V/m
430-470 MHz: 28 V/m
704-787 MHz: 9 V/m
800-960 MHz: 28 V/m
1700-1990 MHz: 28 V/m
2400-2470 MHz: 28 V/m
5100-5800 MHz: 9 V/m
Rated power frequency
magnetic fields
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz