Electromagnetic Conformity
NeoBeat is intended for use in the following environments: Health care
facilities except for near HF surgical equipment and the RF shielded room
for magnetic resonance imaging.
No particular actions are required to maintain safety and performance
with regard to electromagnetic disturbances for the expected service life.
Warnings
• Use of this equipment adjacent to or stacked with other equipment should
be avoided because it could result in improper operation. If such use is
necessary, this equipment and the other equipment should be observed to
verify that they are operating normally.
• Use of accessories, transducers and cables other than those specified or
provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
• Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30
cm (12 inches) to any part of the NeoBeat, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment
could result.