8. Troubleshooting
82
Error Message
Description
Troubleshooting
“Channel Low! Channel:
channel_nr
, Minimum:
minimum
, Maximum
maximum
Signal on channel
channel_nr
did not
reach the expected
area between
minimum
and
maximum
Hardware problem: Electronic
defect, broken fiber
“Invalid Wavelength!
Wavelength:
wavelength
nm”
Filter wavelength is
outside the range of
wavelengths of the
white and the blue LED
System error – report to customer
support
“Lamp Overflow!” Minimum:
minimum
, Maximum:
maximum.
Signal on ADC
exceeds the expected
area between
minimum
and
maximum
Hardware problem: Electronic
defect
“Value Not Set:
value
-1
”
The value
value
is not
set
System error – report to customer
support
“Filter Init Error”
Filter transport could
not be initialized
Hardware problem: Electronic
defect, filter wheel transport
mechanically blocked
8.1.2
Definition of ‘Overflow’
If the result of the absorbance measurement is not within the instrument
specifications (> 4.0 OD) an overflow will occur and the measured OD value of
the respective well will be replaced by ‘Overflow’. This is done by the controlling
software and not by the instrument itself.
8.1.3
Power Failure
In case of power failure the following happens:
Power failure of the instrument, but not of the controlling computer (e.g. computer
connected to uninterruptible power supply system): USB connection between
instrument and computer will be lost. Error message by the controlling software is
created.
Power failure of both instrument and controlling computer: the computer has to
reboot. No measurement data will be available.