5. component description
Test stand housing (A)
Thanks to its ergonomic design, the test stand
can be operated fatigue
-
free even during con-
tinuous operation. Operation in computer mo-
de is via PC keyboard and mouse or manually
via the optionally available front panel
Front cover / Panel (only with Option T) (B)
The front panel contains the control buttons
and an OLED display showing measuring va-
lues and system information..
Multifunction test head (C)
The test head is a fixed head that can be in-
flated by an electric pump so that full face
masks of any brand can be sealed without
problems. A measuring point is installed in the
test head which measures the internal pres-
sure in the mask or demand valve.
The Labtec ECO dynamic model has a se-
cond measuring point in the forehead area of
the head. This can be queried via the soft-
ware.
Air supply / device connection (D)
The air supply can be provided optionally:
1.
Through an existing, fixed 300 bar pipe
supply via a 6 mm high
-
pressure pipe
fitting.
2.
Through an available 300 bar high pres-
sure hose for connection to a breathing
air cylinder or pipe supply.
Attention: Never connect or disconnect
lines under pressure !
Device connection (E)
High
-
pressure device socket for connecting
the pressure regulator of the SCBA to be
tested.
CPS test connection and system vent (F)
Quick coupling for filling CPS. For all other
test procedures, the coupling must be
connected to the supplied plug
-
in silencer for
system venting.
Medium pressure device connections (G,
H)
Quick coupling for connecting the medium
pressure line of the demand valve to be
tested.
Plug nipple for connecting the medium pres-
sure line of SCBA to be tested.
Power connection (I)
On the left side of the housing there is a C13
power plug coupling for the connection of a
cold appliance cable with C14 plug.
USB
-
Interface (J)
The USB port for data transfer to a PC is lo-
cated on the left side of the housing.
Artifical Lung
Labtec ECO dynamic
devices contain the
artificial lung, which is continuously adjustable
in stroke and breathing frequency. The desi-
red respiration can be infinitely adjusted in the
software in the stroke from 5 to 40 strokes/
min and in the volume from 0.5 to 3.5 litres.
Vacuum pump
With the built
-
in vacuum pump, the required
air currents are generated and the probe is
inflated or deflated. The desired air flow rate
can be set in the software from 0.5 to 10 l/
min.