1. foreword

 

 

This operating manual describes the operati-

on of the 

Labtec  ECO 

series of test  instru-

ments.

 

 

 

Labtec  test  devices  are  professional  tools 

that  require  comprehensive  respiratory  pro-

tection expertise, e.g. by completing training 

as  a  respirator  maintenance  technician,  to 

ensure proper operation and correct measu-

rement.

 

 

A prerequisite for proper operation of the de-

vice is that the device is operated and used 

exclusively  in  accordance  with  these  opera-

ting instructions. Labtec is not liable for per-

sonal injury or damage to property or conse-

quential damage of any kind if the device  is 

used  in  a  manner  other  than  that  described 

in this manual. The same applies  to the ac-

ceptance  of  any  warranty  and  guarantee 

claims.

 

 

Otherwise, the terms of sale and delivery as 

well  as  the  warranty  and  guarantee  conditi-

ons  of  Labtec  GmbH  apply,  which  are  not 

extended in any way by these instructions. 

 

 

2. safety instructions

 

 

Each  Labtec  test  device  requires  device

-

specific  instruction  by  an  authorized  Labtec 

employee or training partner.

 

 

The device may only be operated by instruc-

ted and authorized personnel.

 

 

The  device  may  only  be  serviced  and  re-

paired by trained and certified Labtec service 

technicians.  Only  original  spare  parts  and 

accessories must be used.

 

 

The device may only be opened by a Labtec 

service technician or, if expressly requested, 

under instruction by a Labtec employee.

 

 

To  ensure  proper  functioning  of  the  device 

and  thus  a  correct  measurement  result,  the 

device must be serviced at least once a year 

by  a  trained  and  certified  Labtec  service 

technician  and  tested  and  calibrated  by  him 

or her using a Labtec calibration device.

 

 

The  list  of  current,  certified  employees,  trai-

ning  partners  and  service  technicians  is  up-

dated  daily  and  is  available  from  Labtec  or 

on the Labtec homepage.

 

 
 

Labtec devices are partly operated with bre-

athing air in the high pressure range of up to 

400bar. Operating a device under high pres-

sure  requires  the  greatest  care.  Before 

connecting  and  disconnecting  a  test  item  / 

breathing apparatus, the test bench must al-

ways  be  completely  depressurized  and  air 

supply  lines,  especially  the  main  valve, 

closed.

 

 

The device connection must always be che-

cked.  Devices  must  always  be  connected 

completely correctly. When connecting to the 

high pressure device connection, the device 

thread  must  be  completely  screwed  to  the 

internal thread of the device connection.

 

 

Before  switching  off  the  test  device  electri-

cally, the test bench must first be completely 

depressurized  and  vented  and  then  all 

connected  devices  and  any  full  face  masks 

must be removed.

 

 

Never test dirty demand valves or masks on 

the test head. There is a risk of dust particles 

getting  into  the  measuring  system,  which 

can impair the function.

 

 

When the test stand is started up, the mouth 

opening  and  the  instrument  connections 

must remain open, otherwise self

-

calibration 

will  be  incorrect  and  the  measurement  data 

will be incorrect.

 

 

The  test  device  must  be  set  up  horizontally 

and rest well on all support buffers.

 

 

Care must be taken to ensure that the mea-

suring  nozzle  in  the  mouth  opening  in  the 

test  head  is  not  clogged.  In  case  of  false 

measuring  values  in  low  pressure,  the  ope-

ning  can  be  cleaned  with  a  0.2 

-

  0.35  mm 

needle.

 

In case of queries or uncertainties 

 

contact the Labtec customer service!

 

Summary of Contents for ECO static

Page 1: ...ec de User Manual Labtec ECO static dynamic Labtec No Content Page 1 foreword 2 2 safety instructions 2 3 device description 3 4 technical data 4 5 component description 5 6 operating elements 6 7 com...

Page 2: ...ained and certified Labtec service technician and tested and calibrated by him or her using a Labtec calibration device The list of current certified employees trai ning partners and service technicia...

Page 3: ...PC connec ted to test device via swivel arm mounting incl pre installed LabNet Profi test software Module S safety valve test System com ponent and module for automatic medium pressure increase for SC...

Page 4: ...on bar 0 25 0 5 low pressure sensor measuring range precision mbar 50 0 50 0 05 complete device Max permissible operating pressure bar 400 power supply V Hz 230 240 50 60 total power electricity W max...

Page 5: ...linder or pipe supply Attention Never connect or disconnect lines under pressure Device connection E High pressure device socket for connecting the pressure regulator of the SCBA to be tested CPS test...

Page 6: ...switches it off Switch button for 10 l m inhalation Forcing pump 10 l min is set in motion by pressing this button LED shows function Pressing button again switches it off Switch button for CPS infla...

Page 7: ...atus is screwed completely into the inter nal thread of the device connection For tests in the medium pressure range dis connect the medium pressure line of the SCBA at the Euro coupling and connect i...

Page 8: ...C Guideline 2016 42 EG and EMC 2014 30 EU including all changes and addendums to date thereto The above mentioned product meets the following standards and guidelines Nous Vous Confirmons que l appare...

Page 9: ...h head inflation button The inflation time varies from mask type to mask type 1 2 3 Set test vacuum with negative pressure button to setpoint value mbar Switch on stop watch Adjust the test pressure w...

Page 10: ...r a preset time min read the pressure change on the display LP and compare with the setpoint 2 1 5 Completely vent the test device All further tests require high pressure Check correct device connecti...

Page 11: ...ith main valve switch 2 2 7 Completely vent the test device 2 4 1 Connect demand valve in combination with a SCBA base unit to the test device Check correct device connection 2 4 2 Open high pressure...

Page 12: ...t pressure until PA pressure gauge shows the set point Read off the SCBA manometer 3 1 5 Read the high pressure on the display HP and compare it with instrument pressure opt Repeat test steps 3 1 4 an...

Page 13: ...test device 5 3 1 Connect demand valve in combination with a SCBA base unit to the test device Check correct device connection 5 2 1 Open high pressure main valve with main valve switch High pressure...

Page 14: ...ation with a SCBA base unit to the test device Check correct device connection 5 4 2 Open high pressure main valve with main valve switch High pressure flows into the unit 5 4 3 Close high pressure ma...

Page 15: ...des Schalters CSA f llen das Bef llen des Anzugs stoppen when the required tightening internal pressure is reached 7 1 8 Press the CPS inflation switch again to stop fil ling the suit Setting the requ...

Reviews: