9) 

JET SPRAY ASSEMBLY 

The stainless steel spray assembly is a 
double fluid nozzle where the inner nozzle 
develops the liquid jet and the outer the 
compressed air jet.  The compressed air 
atomises the liquid as it emerges from the 
jet to form the required fine spray. 
The jet and outer nozzle can be unscrewed 
from the main assembly for cleaning as 
required. 
Always be careful not to damage the inner 
jet nozzle (which is only 0.5mm 

 thin wall 

stainless steel) as any slight misalignment 
will affect the efficiency of the spray. 
The jet assembly is easy to disassemble by 
unscrewing the knurled rings. 
 
Instructions for testing the jet spray 
assembly outside the main chamber 
 
Before fitting the jet into the top chamber it 
is useful to operate it outside, with water, to 
see the actual spray formation.  First 
connect the silicon tubing to the pump head 
by the following procedure. 
 

 

 

De-blocker inlet (A) 
Sample Inlet (B) 
Atomising air inlet (C ) 

 
Open the hinged plastic cover of the pump and leaving sufficient tube to connect to the jet when it is 
in the top chamber, insert the tube rotate pump head by hand, until positioned under the pump 
rollers.  
 
Open the white plastic clips which retain the tube in position, insert the tube fully and release the 
clips.  The tube should now be regularly compressed by the pump rotors.  Fit one end of the tube to 
the sample inlet (B). 
 
Using the larger 6mm diameter white nylon tube emerging from the top left side of the unit connect 
the free end into the atomising air inlet (C ) on the jet assembly.  This is the main compressed air 
feed line to the jet nozzle.  
 
Then connect the small diameter blue tube into the de-blocker inlet (A).  This is the compressed air 
supply for the jet de-blocking system.  There is no possibility of connecting these tubes incorrectly 
as the connectors are of different sizes.  To release these tubes simply push in the black outer ring 
on the connector and pull the tube out. 
 

Summary of Contents for SD-06

Page 1: ...4 Components Included 5 Packing List 6 Connection to Electrical Supply 7 Assembly and Preparation 8 Controls 9 Jet Spray Assembly 10 Operation 11 General Advice and Useful Hints 12 Optional Accessorie...

Page 2: ...r is the result of more than 30 years of continuous development The SD 06 has been designed primarily for simplicity and ease of use rapid assembly and disassembly for cleaning minimal maintenance eff...

Page 3: ...ble to expose any person to hazardous solids liquids dusts vapours gases or micro organisms You are also required to introduce suitable procedures for handling these substances and to keep appropriate...

Page 4: ...or Residual Current Circuit Breaker RCCB to protect the operator from electric shock in the event of misuse or accident e Potential hazards should always be the first consideration when deciding on a...

Page 5: ...S2023 PTFE Transition Piece 1 SDS2031 Silicone Tube 4 8mm bore x 1 6mm wall 2 0 m SDS2022 Exhaust Tube 1 7 m Nylon Tube 4mm fitted to unit 1 0 m Nylon Tube 6mm fitted to unit 1 0 m The above comprises...

Page 6: ...he U shaped clamp on the left hand side of the unit slacken the two black knobs and slide the main chamber in towards the central control panel Lightly tighten the two black knobs with the 90 outlet o...

Page 7: ...y and the blue nylon tube into the push in fitting at the top of the assembly The silicone tube for the liquid sample is fitted to the grooved inlet The remaining tube is fitted through the pump and t...

Page 8: ...on is used in the set De Blocker screen to change the De Blocker from Slow to Medium or Fast 7 The next button is active on all screens and is used to select the next screen 8 The Stop button is activ...

Page 9: ...lowing procedure De blocker inlet A Sample Inlet B Atomising air inlet C Open the hinged plastic cover of the pump and leaving sufficient tube to connect to the jet when it is in the top chamber inser...

Page 10: ...p setting screen The information indicates pump speed on a nominal 50 scale For appropriate flow rates in ml hr please see the chart on Note II The compressed air pressure setting is factory set to a...

Page 11: ...l times on the set temp screen A fixed over temperature protection cut out is fitted and will actuate at approximately 300 C if actuated allow to cool down and re start allow inlet temperature to stab...

Page 12: ...llow the main chamber a few minutes to heat up The inlet temperature may show the required setting but the glassware and main chamber will be quite cool for a few minutes Wait until inlet temperature...

Page 13: ...15 ml hour 45 RPM 1895 ml hour 40 RPM 1665 ml hour 35 RPM 1500 ml hour 30 RPM 1305 ml hour 25 RPM 1090 ml hour 20 RPM 900 ml hour 15 RPM 695 ml hour 10 RPM 485 ml hour 5 RPM 280 ml hour SD 06 Drying A...

Page 14: ...a logging 13A MAINTENANCE Regular maintenance of any equipment ensures maximum trouble free life Customers are recommended to carry our simple maintenance at regular intervals the most basic of which...

Page 15: ...00mm long SDS2024 Chimney Extension required when using SDS2020 SDS2022 Cyclone SDS2023 PTFE Connection Tube SDS 0021 Clamp for Chamber Cyclone to PTFE Connection Tube SDS2006 Sample Collection Bottle...

Page 16: ...g the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specific...

Reviews: