![LabPlant SD-06 Assembly & Operating Instructions Download Page 14](http://html1.mh-extra.com/html/labplant/sd-06/sd-06_assembly-and-operating-instructions_3197635014.webp)
14
12)
OPTIONAL ACCESSORIES
1) 0.5mm jet is supplied as standard. 1.0mm and 2.0mm jets are optional.
2) Amber stained glassware is available for products which are light sensitive.
3) An exhaust air wet scrubber unit.
4) Magnetic stirrer hotplate MR 3001 for maintaining liquids in suspension and for heating. The
optional side tray is designed to accept this stirrer.
5) Magnetic hotplate MR 3000 when heating is not required.
6) Integral floor stand to give optimum working height when standard bench space is not available.
7) Large main chamber - 800 mm long. Gives added spray drying efficiency but must be used with
optional floor stand. Supplied with a connection tube between the cyclone and the stainless
steel exhaust assembly.
8) Full operational software for remote operation and data logging.
13A) MAINTENANCE
Regular maintenance of any equipment ensures maximum trouble-free life.
Customers are recommended to carry our simple maintenance at regular intervals – the most basic
of which is to keep the equipment clean! No matter how regularly your equipment is in use always
find time to thoroughly clean it.
Periodically vacuuming of the rear air filter and replacement every 12 months will prolong
performance.
Annual service and maintenance contracts are available in the UK
13B) REPLACEMENT CONTROLLER
Control modules for the SD-06 are designed to be exchanged, there are no serviceable parts.
Replacement boards will be despatched on receipt of returned faulty modules
SD-06 PLC
Main P.L.C. module with software.
SD-06 PSU
Power supply unit.
SD-06 HMI
H.M.I. screen with software.
SD-06 PM
Peristaltic pump motor board.
SD-06 A1-2
Analogue unit (state unit 1 or 2).
SD-06 SC
P.R.T. signal conditioning unit.
Please contact your supplier for spare parts pricing information.