LabPlant SD-06 Assembly & Operating Instructions Download Page 8

8) 

CONTROLS 

 
The green indicator (E) on the lower left hand side of the control panel, when illuminated indicates 
that the SD-06 is connected to the electrical supply. 
 
The on-off (I-0) switch (F) on the lower right hand side is used to turn the electrical power to the SD-
06 on & off. 
 
All other control settings and start/stop are effected by means of the microprocessor controller (O). 
 
Study the two diagrams, screen functions & screen details carefully. Familiarise and practise moving 
from screen to screen and changing values (see next pages) 
 

Notes. 
 

1) 

You can only change values on the screen displayed 

2) 

Values can be changed on the following screens :- 

 

a) 

SET TEMPERATURE.  

(0-250) 

next 

b) 

SET FAN 

 

 

(0-50) 

next 

c) 

SET PUMP 

 

 

(0-50) 

next 

d) 

SET DEBLOCKER 

 

(Slow. Medium. Fast) 

 
3) 

The above screens will only allow values within the above range. 

4) 

Pressing ESC (escape) during data input will restore the original value.  

5) 

Start & Pump buttons only operate on the Start – Up screen. 

6) 

The start button is used in the set De-Blocker screen, to change the De-Blocker from Slow to 
Medium or Fast. 

7) 

The next button is active on all screens and is used to select the next screen. 

8) 

The Stop button is active in ALL screens and will immediately  stop the pump and jet De-
Blocker. After 20 seconds delay the fan and compressor will also stop. 

9) 

When the SD-06 is stopped, using the Stop button, all settings are retained. 

10) 

When the Start buttons on the Start Up screen is pressed the fan and compressor will start 
immediately.  The heat does not switch on for 10 seconds. 

 
Remote Control and Data Logging 
 
The SD-06 unit can be remotely controlled by means of a computer and leads via a RS232/485 link 
(found on the reverse of the unit) allowing the process to be controlled and monitored (in real time) 
from a PC. During a monitored run the process conditions can be printed out to give a hard copy 
record. 2 versions of the software are available from Lab-Plant UK Ltd -  the ‘full’ version for 
unlimited use and the ‘limited’ version which allows up to 20 x 1 hour sessions. A set of leads and a 
RS 232/485 converter is required, these can also be obtained from Lab-Plant UK Ltd. 

 

Summary of Contents for SD-06

Page 1: ...4 Components Included 5 Packing List 6 Connection to Electrical Supply 7 Assembly and Preparation 8 Controls 9 Jet Spray Assembly 10 Operation 11 General Advice and Useful Hints 12 Optional Accessorie...

Page 2: ...r is the result of more than 30 years of continuous development The SD 06 has been designed primarily for simplicity and ease of use rapid assembly and disassembly for cleaning minimal maintenance eff...

Page 3: ...ble to expose any person to hazardous solids liquids dusts vapours gases or micro organisms You are also required to introduce suitable procedures for handling these substances and to keep appropriate...

Page 4: ...or Residual Current Circuit Breaker RCCB to protect the operator from electric shock in the event of misuse or accident e Potential hazards should always be the first consideration when deciding on a...

Page 5: ...S2023 PTFE Transition Piece 1 SDS2031 Silicone Tube 4 8mm bore x 1 6mm wall 2 0 m SDS2022 Exhaust Tube 1 7 m Nylon Tube 4mm fitted to unit 1 0 m Nylon Tube 6mm fitted to unit 1 0 m The above comprises...

Page 6: ...he U shaped clamp on the left hand side of the unit slacken the two black knobs and slide the main chamber in towards the central control panel Lightly tighten the two black knobs with the 90 outlet o...

Page 7: ...y and the blue nylon tube into the push in fitting at the top of the assembly The silicone tube for the liquid sample is fitted to the grooved inlet The remaining tube is fitted through the pump and t...

Page 8: ...on is used in the set De Blocker screen to change the De Blocker from Slow to Medium or Fast 7 The next button is active on all screens and is used to select the next screen 8 The Stop button is activ...

Page 9: ...lowing procedure De blocker inlet A Sample Inlet B Atomising air inlet C Open the hinged plastic cover of the pump and leaving sufficient tube to connect to the jet when it is in the top chamber inser...

Page 10: ...p setting screen The information indicates pump speed on a nominal 50 scale For appropriate flow rates in ml hr please see the chart on Note II The compressed air pressure setting is factory set to a...

Page 11: ...l times on the set temp screen A fixed over temperature protection cut out is fitted and will actuate at approximately 300 C if actuated allow to cool down and re start allow inlet temperature to stab...

Page 12: ...llow the main chamber a few minutes to heat up The inlet temperature may show the required setting but the glassware and main chamber will be quite cool for a few minutes Wait until inlet temperature...

Page 13: ...15 ml hour 45 RPM 1895 ml hour 40 RPM 1665 ml hour 35 RPM 1500 ml hour 30 RPM 1305 ml hour 25 RPM 1090 ml hour 20 RPM 900 ml hour 15 RPM 695 ml hour 10 RPM 485 ml hour 5 RPM 280 ml hour SD 06 Drying A...

Page 14: ...a logging 13A MAINTENANCE Regular maintenance of any equipment ensures maximum trouble free life Customers are recommended to carry our simple maintenance at regular intervals the most basic of which...

Page 15: ...00mm long SDS2024 Chimney Extension required when using SDS2020 SDS2022 Cyclone SDS2023 PTFE Connection Tube SDS 0021 Clamp for Chamber Cyclone to PTFE Connection Tube SDS2006 Sample Collection Bottle...

Page 16: ...g the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specific...

Reviews: