background image

 SLx 40 

GUIDANCE TABLES (continued)

The  SLx 40  is intended for use in the electromagnetic environment specified below. The
customer or user of the SLx 40 should ensure that it is used in such an environment.

24

Table 204– Guidance and Manufacturer’s Declaration 

Electromagnetic Immunity

Equipment and Systems that are NOT Life-supporting

Guidance and Manufacturer’s Declaration – Electromagnetic Immunity

Immunity

Test

IEC 60601

Test Level

Compliance

Level

Electromagnetic 

Environment 

- Guidance

Conducted RF 

IEC 61000-4-6

3 vrms

150 kHz to 80 

MHz

(v1) = 3 vrms

Recommended Separation Distance: 

d=(3.5/v1)(Sqrt P) 

d=(3.5/E1)(Sqrt P) 

80 to 800 MHz 

d=(7/E1)(Sqrt P) 

800 MHz to 2.5 GHz 

     

Radiated RF 

IEC 61000-4-3

80 MHz to 2.5 

GHz @ 3V/m

(E1) = 3 

v/m

Note 1:  At 80 MHz and 800 MHz, the higher frequency range applies. 

Note 2:  These guidelines may not apply in all situations.  Electromagnetic propagation is affected by absorption
and reflection from structures objects and people. 

* Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land 

mobile radios, amateur radio, AM and FM radio broadcast and Tv broadcast cannot be predicted theoretically 
with accuracy.  To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic 
site survey should be considered.  The measured field strength in the location in which the ME Equipment or
ME System should be observed to verify normal operation.  If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the ME Equipment or ME System.

* Over the frequency range 150 kHz to 80 MHz, field strengths should be less then [V1] V/m. 

8126400-795 

 SLx 40 

Issue 1.3

Printed on April,2019

Portable and mobile Rf communications
equipment should be used no closer to any
part of the SLx 40, including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the 
transmitter.

Where  P is  the  max  output  power  rating  of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the recom-
mended  separation  distance  in  meters  (m).

Field strengths from fixed Rf transmitters,
as determined by an electromagnetic site
survey, should be less than the compliance
levels in each frequency range.

Interference  may  occur  in  the  vicinity  of
equipment  marked  with  the  following
symbol.

Summary of Contents for SLx 40

Page 1: ...8126400 795 SLx 40 Issue 1 3 Printed onApril 2019 SLx 40 Slit Lamp User Manual...

Page 2: ...Reserves the right to make change in the product described in this manual without notice and without incorporating those changes in any products already sold ISO 13485 Certified LABOMED products are d...

Page 3: ...INSTRUCTIONS FOR USE 12 OPERATION 12 SLIT LENGTH 13 FILTER DIAL 13 SLIT ROTATION 13 CHANGING THE HALOGEN BULB 14 CHANGING THE LED 15 REPLACEMENT OF CHIN REST LIGHT INDICATOR 15 FUSE REPLACEMENT 16 TRO...

Page 4: ...SHOCK THIS EQUIPMENT MUST ONLY BE CONNECTED TO A SUPPLY MAIN WITH PROTECTIVE EARTH OR DAMAGE TO THIS INSTRUMENT AND OR INJURY TO THE OPERATOR OR PATIENT MAY OCCUR WARNING ENSURE THAT THE VOLTAGE APPL...

Page 5: ...MAY OCCUR SEE THE CARE AND MAINTENANCE SECTION FOR DETAILED CLEANING INSTRUCTIONS CAUTION MEDICAL ELECTRONIC EQUIPMENT NEEDS SPECIAL PRECAUTIONS WITH RESPECT TO ELECTROMAGNETIC CHARGE EMC AND NEEDS TO...

Page 6: ...Power Protective Earth Connection ON OFF Manufacture Catalog Number Serial Number Waste of Electrical and Electronic Equipment Compliance to medical Device Directive 93 42 EEC Accompanying Document m...

Page 7: ...and eye care technicians should operate this instrument Please retain this manual for future reference and to share with other users Additional copies can be obtained from your authorized LABOMED dea...

Page 8: ...lumination Arm Lock Knob 17 Microscope Arm Lock Knob 18 Instrument base Lock Knob 19 Guide Rail Cover 20 Geared Rollers 21 Guide Rails 22 Chin Rest Assembly 23 Power Supply Assembly 24 Table Top 25 Pa...

Page 9: ...e User s Guide and follow it for Slit Lamp installation 3 Open the box marked as Platform Assembly and take out the table top and power supply bracket shown as A in fig 1 4 Open the box marked as Chin...

Page 10: ...re 5 8 Install the Slit Lamp Assembly onto the tracks of the Table Top and slide the Guide Rail Covers around the tracks Refer to Figure 6 9 Assemble Power Supply to the table top just by pushing into...

Page 11: ...ION OF INPUT POWER WARNING CARE MUST BE TAKEN TO ARRANGE THE CABLES FOR THE ACCESSORIES SUCH THAT THEY DO NOT PRESENT A TRIPPING HAZARD TO THE EXAMINER OR A DANGER TO THE PATIENT WARNING POSITION THIS...

Page 12: ...allow for greater accuracy 5 Remove the Focusing Rod 6 To position a patient adjust the chinrest height by turning the Chinrest Elevation Handle on the post of the Chin Rest Assembly until the patient...

Page 13: ...he Magnification Dial on the Microscope Assembly Refer to figure 16 SLIT LENGTH The Slit length is adjusted by rotating the Slit Length Dial The dial has five stops for adjustments They are 0 6 5 8 9...

Page 14: ...B WITH BARE HANDS AS FINGERPRINTS WILL SHORTEN THE BULB LIFE 1 Remove input power to the instrument 2 Open the bulb door 3 Swing the retaining spring away from tha Bulb Refer to figure 21 4 Pull the B...

Page 15: ...by using circlip plier Refer Figure 24 3 Lift up the illumination housing carefully 4 Remove the Lamp House Cover by unscrewing the Four Screws by using screwdriver shown as B refer in Figure 25 5 Un...

Page 16: ...n on the top tab in the middle of the Power Input Module to release the Fuse Holder and gently pull out the Fuse Holder by gripping the two small tabs Refer to figures 24 25 3 Open the Door to the Fus...

Page 17: ...ore use Install focusing rod and check to ensure microscope is focused on it Bulb not installed properly Finger Prints on CMO Illumination lamp is not fitted in place Aperture slit length selector tur...

Page 18: ...106 kPa 31 3 in Hg Transportation Storage 10 Temperature 20 C 4 F to 70 C 158 F 106 kPa Relative Humidity 10 to 80 non condensing Atmospheric Pressure 50kPa 14 8 in Hg to 106 kPa 31 3 in Hg 50 kPa Ex...

Page 19: ...quipment to a protective grounding system so that the accessi ble metal parts do not become conductive in case of failure in the basic insulation Grade of protection against a hazardous ingress of wat...

Page 20: ...nic reasons wipe the forehead rest with an alcohol wipe and change the chin rest papers after each patient CLEANING THE GUIDE PLATE If the Guide Plate is dirty it may cause a rough feeling when maneuv...

Page 21: ...T CCD ADAPTER 8122500 8122600 800 6134131 6134110 6134120 6134172 I PHONE 6 6134172 1 6 I PHONE 6PLUS 6134172 1 6P 8122600 851 8122600 852 CCD CAMERA I PHONE 7 6134172 1 7 SONY HANDICAM 6134120 122 I...

Page 22: ...Test Compliance Electromagnetic Environment Guidance RF Emissions CISPR 11 Group 1 Class A Harmonics IEC61000 3 2 Class A Flicker IEC61000 3 3 Complies 8126400 795 SLx 40 Issue 1 3 Printed onApril 201...

Page 23: ...5 Cycle 60 Dip for 5 Cycles 30 Dip for 25 Cycles 95 Dip for 5 Seconds 95 Dip for 0 5 Cycle 60 Dip for 5 Cycles 30 Dip for 25 Cycles 95 Dip for 5 Seconds Power Frequency 50 60Hz Magnetic Field IEC61000...

Page 24: ...eur radio AM and FM radio broadcast and Tv broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site surv...

Page 25: ...mmended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts w ac...

Page 26: ...ENT WARRANTY If notified promptly on writing of any action brought against the purchaser based on a claim that the instrument infringes a U S Patent Labo America Inc will defend such action at its exp...

Page 27: ...Revision History Rev No Date of Release Change App By DCR 1 3 April 4 2019 System Diagram added S Bal DCR 11 19...

Page 28: ...hone 510 445 1257 Fax 510 991 9862 Email sales laboamerica com www laboamerica com Labomed Europe Essebaan 50 NL 2908 LK Capelle a d IJssel The Netherlands Tel 31 0 10 4584222 Fax 31 0 10 4508251 E ma...

Reviews: