Guidance and Manufacturer’s Declaration
Electromagnetic Immunity
All Equipment and Systems
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The Prima GN is intended for use in electromagnetic environment specified below. The
customer or user of the Prima GN should ensure that it is used in such an environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment
- Guidance
ESD
IEC 61000-4-2
±6kv Contact
±8kv Air
±6kv Contact
±8kv Air
Floors should be wood, concrete or ceramic
tile. If floors are synthetic, the R/H should be at
least 30%.
EFT
IEC 61000-4-4
±2kv Mains
±1kv I/Os
±2kv Mains
±1kv I/Os
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±1kv Differential
±2kv Common
±1kv Differential
±2kv Common
Mains power quality should be that of a typical
commercial or hospital environment.
voltage
Dips/Dropout
IEC 61000-4-11
>95% Dip
for 0.5 Cycle
60% Dip
for 5 Cycles
30% Dip
for 25 Cycles
>95% Dip
for 5 Seconds
>95% Dip
for 0.5 Cycle
60% Dip
for 5 Cycles
30% Dip
for 25 Cycles
>95% Dip
for 5 Seconds
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the Prima GN requires continued operation
during power mains interruptions, it is recom-
mended that the Prima GN be powered from
an uninterruptible power supply or battery.
Power Frequency
50/60Hz
Magnetic Field
IEC 61000-4-8
3A/m
3A/m
Power frequency magnetic fields should be that
of a typical commercial or hospital environment.
2
9
6165000-795
PRIMA GN