16
6 Care & Maintenance
•
Maintenance
1. Headrest/Chinrest Cleaning
For hygienic reasons, wipe the headrest and chin rest with an alcohol, wipe after every patient.
2. Changing Illumination Lamp
Ensure the Slit Lamp is switched off and remember that some parts may be hot. Do not
touch the glass of the new bulb.
Open the bulb access door. Swing the retaining spring towards the arm pivot and pull
the socket and bulb from the housing. Note the position of the notch in the metal collar
of the bulb. Replace the bulb with a new one and make sure the notch is in the same
position. Insert the bulb and the socket into the lamp housing and move the retaining
spring back into position. Close the bulb access door.
3. Projection Pros, Cleaning
Grasp the narrow shank of the projection prism and pull upwards. When cleaning the
prism, blow with clean dry air, then, gently wipe with a soft linen cloth.
4. Cleaning the Slide Plate
If the plate is dirty, it may cause a rough feeling when maneuvering the base of the
Slit Lamp. Clean the slide plate with soft cloth lightly dampened with a mild soap and
water solution.
CONTINUOUS POWER SUPPLY
The LABOMED eVO500/eVO500D Slit Lamp can be operated from 100 to 240V AC. This
is achieved by universal power supply. So, there is no need of adjusting the settings
according to the supply line voltage of the destination country.