4   Operating Procedure

 

 

1.

 

Turn the power using the ON/OFF switch located on the front of the power supply. 

Brightness can be adjusted by rotating the illumination level knob situated on the 

microscope base. 

NOTE: The maximum position is for intermittent use only.  Continuous use will 

shorten lamp life. 

2.

 

Insert the projection rod in the pivot post of the instrument body to make rough PD 

and focus adjustments. Position the light onto the flat surface of the projection rod 

and adjust the pupillary distance and focus of the eyepieces till a comfortable 

viewing is obtained. Remove the projection rod. 

3.

 

To position a patient, adjust the chinrest height by turning the control knob on the 

post of the chin rest assembly until the patient’s canthus is in line with the canthus 

mark on the chin rest post. 

4.

 

Microscope elevation is adjusted by rotating the joystick and observing the Slit 

image through the microscope until the slit is centered on the patient’s cornea. 

5.

 

Move the Slit lamp with the joystick held firmly and slightly angled towards the 

patient, until the slit appears sharply on the cornea. The accuracy of this rough 

adjustment should be checked by the naked eye. The fine adjustment is performed 

while observing the slit through the microscope. 

6.

 

Tilt the joystick, which is now held lightly at its upper end, until the slit appears 

sharply at the depth of the eye which is to be observed. The horizontal motion of 

the base can be locked by tightening the base locking screw. Lock the base 

whenever the slit lamp is not in use. 

7.

 

The slit width can be adjusted by rotating the slit width control knob on either sides 

of the instrument. 

8.

 

The angle between the illumination system and the microscope can be varied 

between 0° to 90° to either the left or to the right. The angle is indicated on the 

scale of the slit lamp arm. 

Summary of Contents for eVO 500

Page 1: ...Digital Slit Lamp eVO 500 500D ...

Page 2: ...1 Contents 1 Introduction 6 7 2 Safety Instruction 8 3 Installation 9 11 4 Operating Procedure 12 13 5 Lamp and Fuse Replacement 14 15 6 Care maintenance 16 18 7 Technical Specification 19 ...

Page 3: ...2 eVO500D ...

Page 4: ... Locking Knob 8 Illumination arm locking arm 9 Lamp access door 10 Filter Dial 11 Slit Length Dial 12 Slit width rotation knob 13 Yellow Filter 14 Magnification knob 15 Focusing rod 16 Back Light 17 Target Light 18 Binocular Head Locking Knob 19 Eyepiece 20 SD card slot 21 Monitor 22 Inlet for Lamp Supply 23 AC inlet 24 Inlet for Back light 25 Fuse Holder ...

Page 5: ...4 eVO500 ...

Page 6: ...ver 6 Instrument Base Locking Knob 7 Microscope arm Locking Knob 8 Illumination arm locking arm 9 Lamp access door 10 Filter Dial 11 Slit Length Dial 12 Slit width rotation knob 13 Yellow Filter 14 Magnification knob 15 Focusing rod 16 Back Light 17 Target Light 18 Binocular Head Locking Knob 19 Eyepiece ...

Page 7: ...hand The excellent transmission of the optical system minimizes light loss during therapy Together with the excellent image quality this provides ideal conditions for crisp crystal clear images of even the finest structure Considering standard work procedures the design of this system have been modified to carry an in builtcamera The integrated module is optically corrected unit to deliver the bes...

Page 8: ... manipulation on the eye eVO500 eVO500D is a powerful instrument with 5 step magnification 6 5x 10x 16x 25x 40x in combination with a binocular straight head 12 5x 16mm wide field eyepieces Diopter adjustment 6 and a smooth adjustable IPD 48 5 80mm An eVO500D slit lamp is designed for all digital imaging requirements Its fatigueless examination with integrated 5 0 mega pixels 1080p digital module ...

Page 9: ...priate for one another If any of the connectors are damaged please contact LABOMED representative This instrument is meant to be used in dry rooms only Care must be taken to ensure that no product containing liquid be kept on the instrument at any time to avoid spillage and seepage of liquids inside the instrument The manufacturer will not accept any liability for damage caused by unauthorized per...

Page 10: ... base platform with the socket head screws hexagonal wrench provided as shown in the Fig 2 3 Plug the wire from the chin rest assembly into the 5 pin connector on the back of the power supply unit Fig 2 4 Mount the base platform on appropriate instrument stand arm or table base 5 Position the Slit Lamp body gently on the base platform so that the geared rollers match with the teeth on the guide ra...

Page 11: ...o position making sure it is up against the stop Then tighten lock knob located on the top of the microscope Refer Fig 5 Fig 5 8 Plug the microscope illumination cable with 4 pin adaptor into the connector on the back of the power supply as well as on the back side of the microscope base Refer Fig 6 9 Connect the 3 pin inlet power cord to power supply and plug into a wall outlet or the motorized t...

Page 12: ...n panel on the microscope arm along with the knob in the mount Screw the mount to the knob to hold it tight incase of eVO500D See Fig 7 11 Connect the USB output at the back of the panel and the power cord incase of eVO500D Fig 7 ...

Page 13: ... elevation is adjusted by rotating the joystick and observing the Slit image through the microscope until the slit is centered on the patient s cornea 5 Move the Slit lamp with the joystick held firmly and slightly angled towards the patient until the slit appears sharply on the cornea The accuracy of this rough adjustment should be checked by the naked eye The fine adjustment is performed while o...

Page 14: ...rior over its basic version eVO500 1 Mount the touch screen panel that allows the user to view live images Use click button provided on the joystick to capture images and record videos 2 Select the options on the touch screen panel as per your requirements These options are Capture Zoom In Out Exit A Capture option once selected will show Capture image and Capture Video this works only if the SD c...

Page 15: ...en the lamp house by loosen the screw 1 See figure 8 1 Fig 8 2 Take out the bulb holder 1 by pulling the metal pins on the holder see fig 9 1 Fig 9 3 Loosen the screw and pull the bulb out of the house assembly Refer Fig 10 1 Fig 10 ...

Page 16: ...nsert the holder in the bulb slowly and tighten the lamp cover after putting it in its place Replacement of Fuse 6 Open the fuse holder provided on the back side of the power supply See fig 12 7 One spare fuse is being provided in the box Replace the dead fuse with the live one Insert the fuse holder back to it s position Fig 11 Fig 12 Fig 13 ...

Page 17: ...nsert the bulb and the socket into the lamp housing and move the retaining spring back into position Close the bulb access door 3 Projection Pros Cleaning Grasp the narrow shank of the projection prism and pull upwards When cleaning the prism blow with clean dry air then gently wipe with a soft linen cloth 4 Cleaning the Slide Plate If the plate is dirty it may cause a rough feeling when maneuveri...

Page 18: ...re chemicals close to the instrument o Protect it from improper handling Never install other devices sockets or unscrew optical system and mechanical parts unless explicitly instructed to do so in this manual o Protect the microscope from oil and grease Never oil or grease the guide surfaces or mechanical parts o Remove coarse contamination using a damp disposable cloth o Use disinfectants based o...

Page 19: ... for processing contaminated products Current regulations of hospital hygiene and prevention of infection must be observed in the preparation cleaning and disinfection of the products Instructions Workplace Remove surface contamination with a paper towel Reprocessing Recommended reprocess a product immediately after use Cleaning Needed water detergent spirits microfiber cloth o Flush the surface w...

Page 20: ...rtures 0 5 5 8 12mm 1 5 11mm continuous slit measurement 0 5 5 8 12mm 1 5 11mm continuous slit measurement Slit Rotation 0 180 0 180 Filters Red Free Cobalt Blue Heat absorbing Red Free Cobalt Blue Heat absorbing Longitudinal Movement 90mm 90mm Lateral Movement 100mm 100mm Vertical Movement 30mm 30mm Chin Rest Range 80mm 80mm Illumination 6V 20W Halogen 6V 20W Halogen Digital Imaging Integrated 5 ...

Page 21: ...20 ...

Page 22: ......

Reviews: