![Labkotec LOKASET R Installation And Operation Instructions Manual Download Page 7](http://html1.mh-extra.com/html/labkotec/lokaset-r/lokaset-r_installation-and-operation-instructions-manual_3190686007.webp)
7
LOKASET R
| Installation and Operation Instructions | D18067Ae
3
5
7
1
2
4
Connect the probe cable conductors to the screw terminals
(5) on the transmitter circuit board. The polarity of the cable
connection has no signifi cance to device operation. Tighten the
probe cable gland (6). Install a cable tie above the cable gland
as a cable clamp. Place the supplied battery CR123A in the
battery compartment (7) with the positive pole (+) facing
upward. Correct installation of the probe cable and battery
is illustrated on page 9.
Ensure that the tightening O-ring (8), Ø 69.5 x 3.0 mm NBR70,
is seated properly in the groove. Close the protective cover (9)
with the supplied stainless steel screws (10); 2 pcs, 3 x 16 mm.
Note! Proper seating of the O-ring and tightness of the cable
gland are important in ensuring the tightness of the transmitter
enclosure. If the cover cannot be properly secured, ensure
that the O-ring is installed correctly.
The Lokaset R transmitter is supplied with a 3 metre cable.
The height of the probe may be adjusted using the provided
cable tie. Make a loop in the probe cable. Secure the loop with
the cable tie (11). This facilitates height adjustment of the probe.
Do not leave the probe in the tank during vacuum disposal. Do
not secure the probe cable to tank walls.