Product Service 1-800-522-7658
155
4.
For 3-foot wide models only – Set the prefilter into the towel catch, ensuring
the filter is captured by the long rails on the top and bottom of the towel catch.
Go to Step #6.
5.
For all other models - To install the kit, press bolt (6) through the towel catch
frame as shown, (Note: the long flanges on the top and bottom of the towel
catch go in the same direction as the bolt. Secure the bolt with lock washer (4)
and hex nut (5), and tighten. Set the prefilters in the towel catch, so they are
captured by the long flanges. Install the bracket (2), and secure it with lock
washer (4) and threaded handle (3).
6.
Reinstall the towel catch/prefilter assembly by sliding it back into position,
ensuring that the long flanges on the top and bottom are pointing towards the
front of the cabinet. It must be positioned as shown in the first illustration. Also
ensure that the black sampling/decontamination tube is secure against the back
of the cabinet bottom.