Product Service 1-800-522-7658
114
Motor/Blower Replacement
Caution: The cabinet must be properly decontaminated before servicing the
motor/blower.
The cabinet blower motor's bearings are sealed and require no lubrication.
DO NOT attempt to lubricate them.
To replace the Motor/Blower:
1.
Unplug the cabinet.
2.
Remove the front dress panel as shown in Figure 6-1.
3.
Loosen all of the Phillips screws on the blower chamber cover. Remove all of the
screws except two screws on the top center of the panel. This will support the
panel.
4.
Release the damper seal by pushing the outside edge of the seal towards the
bolt and inside the blower plenum cover, as shown in Figures 6-14 and 6-15.
5.
Remove the last two screws. Gently rock the cover, slide it to the left or right,
and swing it out and away from the unit.
NOTE: During disassembly, the blower plenum cover can rest on the central sash
support, as shown in Figure 6-16, to allow for repositioning of the panel or your grip on
it.
6.
Disconnect both motor connectors from the motor. The 5-pin power connector,
located closest to the blower, is released by pressing the locking ears on either
side of the connector. The 16-pin data connector, located furthest from the
blower, is released by pressing the single locking ear located on the inside center
surface of the connector body. The connectors are shown in Figure 6-28.
Figure 6-28
ECM Power Connector Release
Tabs
ECM Data Connector Release Tab