Confirming Calibration Procedure-
With the cabinet operating:
1.
Press the
Menu
button.
2.
Press the
▼
button until the
Services
option is highlighted
(it will turn yellow).
3.
Press
OK/Mute
to enter the Services menu screen.
4.
Press the
▼
button until the
Diagnostics
option is
highlighted (it will turn white).
5.
Press
OK/Mute
to enter the Diagnostics submenu screen.
6.
Press the
▼
button until the
Airflow Sensor
option is
highlighted (it will turn white).
7.
Press
OK/Mute
to enter the Airflow Sensor submenu
screen.
8.
The display shows the following values in millivolts -
a.
“Zero” – should be approximately 90-110.
b.
Alarm (approximately 85% of Full scale – calculated
by the cabinet)
c.
Full scale - the value for inflow set during
calibration.
NOTE: There should be approximately 100 millivolts or higher difference
between the zero and operational (full scale) points. Differences
significantly lower than 100 mV may result in erratic air velocity values
being displayed
.