71
Chapter 8: Troubleshooting
PROBLEM
CAUSE
CORRECTIVE ACTION
Airflow Monitor
Malfunction
(Continued)
Wrong alarm set
point.
Airflow monitor was not properly
adjusted. Repeat calibration steps
outlined in this manual in Chapter 6
or in the supplied Digital 1000
Monitor Manual.
Constant
audible
alarm.
Check airflow and calibration of
airflow monitor. See Chapter 6.
Continuous alarm.
Check the face velocity of the
enclosure as the airflow of the
system may have changed. If
incorrect, adjust the speed control to
increase face velocity. The HEPA
filter may have become loaded. If
face velocity is correct, calibrate the
airflow monitor outlined in this
manual in Chapter 6.
Monitor alarms; air
way to airflow
monitor sensor is
blocked by insects,
dust or debris.
Lightly clean the airway with clean
air. Be careful not to touch sensitive
electrical components.
Audible
disable
will not stay
operational.
An alarm condition must be
continuously present before the
audible alarm can be silenced. If
flow conditions fluctuate near the
alarm set point, the airflow monitor
will automatically reset it. Action
should be taken to bring the
enclosure airflow into proper
operating parameters or adjust the
alarm set point lower.
Summary of Contents for 3940200
Page 2: ......
Page 19: ...14 Chapter 3 Getting Started Figure 3 1 Filtered Enclosure Installation...
Page 73: ...68 Chapter 7 Accessorizing and Modifying your Filtered Enclosure...
Page 85: ...80 Appendix B Dimensions and Exhaust Options Figure B 1 XPert Filtered Balance System...
Page 86: ...81 Appendix B Dimensions and Exhaust Options Figure B 2 XPert Filtered Balance Station...
Page 90: ...85 Appendix C Filtered Enclosure Specifications C 1...
Page 91: ...86 Appendix C Filtered Enclosure Specifications C 2...
Page 92: ...87 Appendix C Filtered Enclosure Specifications C 3...
Page 93: ...88 Appendix C Filtered Enclosure Specifications C 4...
Page 100: ...95...
Page 101: ......