28
Chapter 4: High Performance Features and Safety
3.
Upper Dilution Air Supply
provides bypass air from
above the work area. This feature constantly bathes the
inside of the sash with clean air and reduces powders,
particulate materials and chemical fumes along the sash
plane, near the critical breathing zone. Five to seven
percent of the required air volume is introduced through the
upper dilution air supply. The upper dilution air supply
also reduces stagnant pockets of air in the upper interior.
See Figure 4-2.
4.
Zoned Rear Perforated Baffle
directs horizontal laminar
air streams to the three-zoned sections of the perforated
baffle. The three-zoned sections have increasingly more
open area at the bottom that help form laminar airflow.
This minimizes the potential for air to roll forward
preventing contaminants from moving toward the sash
opening. The majority of contaminants are highly diluted,
captured and removed on the first pass through the
enclosure. See Figures 4-1 and 4-2.
5.
Side-Entry Air Foils
allow turbulence-free air to enter the
enclosure from the sides and allow clean air to sweep the
interior walls of the enclosure. See Figures 4-1 and 4-2.
6.
Ergonomic Slope
of 10 degrees provides maximum
visibility, and comfort reduces glare, thereby minimizing
operator fatigue. See Figure 4-1.
7.
Internal Depth of 23"
provides necessary depth to support
modern balances and other auxiliary equipment without
extending outside the enclosure or resting on the lower air
foil. See Figure 4-1
8.
Safety Glass Sash with Spring-Loaded Latch
has a
wiping seal to contain contaminants and features a spring-
loaded latch to secure sash open for loading and cleaning.
The sash must be down for normal operation. See Figure
4-1.
9.
Utility Ports with Iris
allows electrical cords and data
cords to pass through the back of enclosure without leaving
a large hole for contaminants to escape. The enclosure
ships with solid plugs. Iris plugs are included with the
User’s Manual. See Figure 4-1 and 4-2.
10.
Accessory Guardian Airflow Monitor
or
Guardian™
1000 Digital Airflow Monitor
continuously monitors
airflow. An audio/visual alarm alerts the user to low
airflow conditions. The Guardian 1000 Digital Airflow
Summary of Contents for 3940200
Page 2: ......
Page 19: ...14 Chapter 3 Getting Started Figure 3 1 Filtered Enclosure Installation...
Page 73: ...68 Chapter 7 Accessorizing and Modifying your Filtered Enclosure...
Page 85: ...80 Appendix B Dimensions and Exhaust Options Figure B 1 XPert Filtered Balance System...
Page 86: ...81 Appendix B Dimensions and Exhaust Options Figure B 2 XPert Filtered Balance Station...
Page 90: ...85 Appendix C Filtered Enclosure Specifications C 1...
Page 91: ...86 Appendix C Filtered Enclosure Specifications C 2...
Page 92: ...87 Appendix C Filtered Enclosure Specifications C 3...
Page 93: ...88 Appendix C Filtered Enclosure Specifications C 4...
Page 100: ...95...
Page 101: ......