7
Before the installation, carefully examine the delivery for possible damage or missing parts. Open the crate and
take the machine out of the box. Check that the machine has not been visibly damaged during the transport.
Please keep the packing material for the duration of the guarantee period. In case of a complaint and return of the
machine in unsuitable packing material, your guarantee claim will be lost.
Check that the mains cord is compatible with the local standard.
If any kind of damage occurred during transport, immediately make a complaint to the carrier. Any incorrect
delivery or missing parts should be reported to the distributor.
4
Installation
4.1.
Unpacking
4.2
Selecting the right place
When selecting the right place for the machine, please consider the following:
-
Put the device on smooth, horizontal and stable place.
-
Leave enough space beyond the device for normal air circulation, min. 200 mm at the back of the machine.
-
Leave enough space around the device, that you will easy control and maintain it.
-
Don
’
t use the device in surroundings, where there are fast temperature and humidity changes. Also avoid places
exposed to direct sunlight and places nearby heating devices.
-
Avoid places, where the possibility of shocks and vibrations exists.
Note: The machine should not be placed so, that it is difficult to pull out the cord plug from mains power
supply.
4.3
Connectiong the power cord
The correct voltage and frequency for the LM250/LM1000 are given on the name plate. Ensure that these values
correspond to the available power supply system.
To avoid interference from noise, surges and spikes, a dedicated line is preferred. If no such line is available,
avoid lines to which powerful electric motors, refrigerators and similar devices are connected.
The power can be turned on and off by the emergency switch, located on the front cover at the right side of the
machine. Light in the switch indicates, when the power is on.
Note that when power is lost the saved programs will be lost!
4.4
Environmental conditions
The machine has been built for operating in laboratory environment. Therefore the environmental conditions
should be the following:
-
Temperature from +5
°
C to +40
°
C
-
Humidity up to 85% RH, non
-
condensing
LM250/LM1000 PULVERISER RING MILL MANUAL