5-22
WinEP600 and SetAddEP600 operating instructions
Normally a fast field measurement is required for faster response and
better stability when the field probe controls as feedback for a preset field
value the radiated power of a system composed by antenna and RF
generator.
However, as faster the measurements as higher the noise is, with
consequent reduction of sensitivity.
Slower measurements increase sensitivity but increase the response time
and the power consumption.
The PMM EP-600/601/602/603/604 features eight different filters to help
finding the most suitable combination of the a.m. parameters for each
specific application:
F
Settling time
Power consumption
Sensitivity
Rejection
@ 50Hz
Rejection
@ 60Hz
1
Very fast
Very low
Low
-----
low
2
Very fast
Very low
Average
Low
------
3
Fast
Low
Good
------
-------
4
Average
Average
High
Good
Very high
5
Average
Average
High
High
High
6
Average
Average
High
Very high
Good
7
Slow
High
Very high
good
Good
8
Slow
Very high
Very high
High
High
In normal operation setting the filters F4 - F5 may offer a good
compromise in terms of power consumption, sensitivity, settling time
and rejection @ 50Hz.
Summary of Contents for narda PMM EP-600
Page 14: ...XIV Safety Consideration This page has been left blank intentionally ...
Page 45: ...Measurements 3 7 Fig 3 3 PMM EP 600 601 602 603 604 in Anechoic Chamber ...
Page 96: ...6 6 Communication protocol This page has been left blank intentionally ...
Page 104: ...8 2 Accessories This page has been left blank intentionally ...
Page 106: ...8 4 Accessories This page has been left blank intentionally ...
Page 108: ...8 6 Accessories This page has been left blank intentionally ...
Page 112: ...8 10 Accessories This page has been left blank intentionally ...
Page 114: ...8 12 Accessories This page has been left blank intentionally ...
Page 116: ...8 14 Accessories This page has been left blank intentionally ...
Page 118: ...Suggerimenti Commenti Note Suggestions Comments Note ...