WinEP600 and SetAddEP600 operating instructions
5-13
5.5.5 XYZ / TOT
Button for selecting the x - y – z (
XYZ
) or Total mode (
TOT
).
Max. display resolution: 0.001 V/m.
Click on the XYZ button to toggle to Total mode.
If the field value measured is outside the nominal level range, the following
messages are displayed:
Ovr :
the field level is higher than 110% of the nominal maximum level (550
V/m for EP601).
! :
the field level is between 100% and 110% of the nominal maximum level
(from 500 to 550 V/m for EP601). The symbol is close to the value.
* :
the field level is lower than nominal minimum level. The symbol is close
to the value.
•
0.5 V/m per EP601 on the total;
•
0.5 /
√
¯ 3 per EP601 on the single axis.
Low :
the field level is lower than 1/15 of the nominal minimum level.
•
0.5 / 15 per EP601 on the total
•
0.5 / (15 x
√
¯ 3) per EP601 on the single axis.
The field value outside the nominal level range could be unreliable.
Summary of Contents for narda PMM EP-600
Page 14: ...XIV Safety Consideration This page has been left blank intentionally ...
Page 45: ...Measurements 3 7 Fig 3 3 PMM EP 600 601 602 603 604 in Anechoic Chamber ...
Page 96: ...6 6 Communication protocol This page has been left blank intentionally ...
Page 104: ...8 2 Accessories This page has been left blank intentionally ...
Page 106: ...8 4 Accessories This page has been left blank intentionally ...
Page 108: ...8 6 Accessories This page has been left blank intentionally ...
Page 112: ...8 10 Accessories This page has been left blank intentionally ...
Page 114: ...8 12 Accessories This page has been left blank intentionally ...
Page 116: ...8 14 Accessories This page has been left blank intentionally ...
Page 118: ...Suggerimenti Commenti Note Suggestions Comments Note ...