1-4
General Information
1.5 HP-01 Main
specifications
When not differently specified the following specifications are referred to an
operating ambient temperature of 23°C and relative humidity 50%.
Table 1-1 Technical specifications of the HP-01 Magnetometer - Field Analyzer
Frequency range
DC ÷ 1000 Hz
Measurement range
(1)
10 µT ÷ 60 mT
(lower range)
10 mT ÷ 10 T
(higher range)
10 µT ÷ 10 T
(advanced autoranging feature)
Overload
20 T
Dynamic range
(after zeroing)
120 dB
Resolution
(2)
100 nT
(lower range)
100 µT
(higher range)
Displayed average noise level
(3)
10 µT
Frequency flatness
(typ)
0.5 dB
(DC ÷ 1000 Hz)
Expanded Measurement Uncertainty
(4)
1%
Field sensor
Triaxial orthogonal Hall effect devices
(temperature compensated)
Sensor area approx.: 0.6 x 0.6 mm
(higher range)
0.15 x 0.15 mm
(lower range)
Acquisition method
Simultaneous triaxial acquisition
Spectrum analysis method
Realtime Gapless
(5)
FFT
SPAN
20 Hz
30 Hz
100 Hz
1000 Hz
Frequency resolution
0.2 Hz
0.3 Hz
1 Hz
3 Hz
RBW
(-3 dB)
0.8 Hz
1.2 Hz
4 Hz
10 Hz
Reading update time
< 350 ms
Start frequency
User selectable within the whole range
Stop frequency
User selectable within the whole range
Time domain analysis method
DC to 1 Hz FIR filter 1 Hz (-3dB), reading updating time 67 ms
Calibration
(6)
internal E
2
PROM
Temperature and humidity meter
-20 to +55 °C 0.01° resolution
5 to 95 % 0.1% resolution
Typical temperature deviation
@ 0 Hz referred to 23°C
0.05% / °C
Dimensions
Sensor stick
φ
9 mm, L 143 mm
Main body
φ
43 mm, L 55 mm
Overall length
200 mm
Weight
100 g
Tripod support
Threaded insert ¼”
Internal battery
3.7 V / 250 mAh Li-Ion, rechargeable
Operating time
> 6 hours
Recharging time
< 2.5 hours
External supply
10 ÷ 15 VDC, I = approx. 100 mA
Optical fiber connection
Serial Optical Interface 115200 Baud
(7)
- RP02 connector
up to 40 m (USB-OC)
Firmware updating
Through the optical link by USB or RS232
Self test
Automatic at power on
Operating temperature
-20 to +55 °C
Operating relative humidity
(8)
5 to 95 %
Charging temperature
0 to +40°C
Storage temperature
-30 to +75°C
(1)
For each single axis. Positive and negative at DC.
(2)
At 20 Hz and 30 Hz SPAN.
(3)
After zeroing. DANL is frequency and SPAN depending. The specified performance is referred to 0 Hz and SPAN 20 Hz.
(4)
At DC.
(5)
Overlapping for SPAN lower than 1000 Hz
(6)
Recommended re-calibration interval 24 months
(7)
38400 Baud with 8053B
(8)
Without condensation
Specifications are subject to change without notice
Summary of Contents for HP-01
Page 14: ...1 6 General Information This page has been intentionally left blank ...
Page 21: ...HP01 TS software 3 3 Fig 3 4 Installation folder Fig 3 5 Ready to install ...
Page 38: ...4 16 Description This page has been intentionally left blank ...
Page 42: ...5 4 Firmware Update This page has been intentionally left blank ...
Page 50: ...8 4 Accessories This page has been intentionally left blank ...
Page 64: ...9 14 Communication protocol This page has been intentionally left blank ...
Page 66: ...Suggerimenti Commenti Note Suggestions Comments Note ...