3-3-2
Connecting Potentiometer Sensor
Measuring Method
Connection Method
Measuring
Channel Mode
•
Potentiometer sensor
Potentiometer
resistance
1to 10 kOHM
Shield
POT.
3-3-3
Connecting Voltage and Current Output Type Transducer
Measuring Method
Connection Method
Measuring
Channel Mode
•
Voltage
measurement
Power source
Shield
V/500 mA
V/50 V
•
Current measurement
R =250 OHM
Highly accurate
resistor
Power source
Shield
I/50 mA
MEMO
Recommended high accuracy resistor for measuring electric current
When it is used with current ranging from 0 to ±50 mA : VSR4 250R00 ±0.02 % (VISHAY made)
When it is used with current ranging from 0 to ±30 mA : VSH1 250RTC (VISHAY made)
- 3-13 -