3-1-4
Connecting DC Power Supply
DC power supply version of the UCAM-65A requires 10 to 16 VDC.
Before connecting the power cable, do not forget to confirm the following.
(1) Confirm that the DC power supply voltage has 10 to 16 VDC.
If the UCAM-65A is used with power other than the above, it may lower the performance and cause
trouble.
(2) Ground the “GND” terminal on the UCAM-65A rear face.
(3) Take special care for the polarity and correctly connect the DC power cable (P-57) to the DC power connector.
MEMO
MEMO
• The minus (‘-’) side of the UCAM-65A DC power connector is connected to the ground. Therefore, do
not connect the UCAM-65A to measuring instruments with plus (‘+’) sides connected to the ground or
do not let the UCAM-65A contact or connect the plus (‘+’) side of the power supply connector.
Or, performance of the UCAM-65A is lowered and cause trouble.
Measuring instrument with ‘+’side connected to the ground
• Do not connect to the minus (‘-’) side of the UCAM-65A DC power connector (=GND or DCOM)
from halfway potential of power voltage.
Other measuring instruments, etc.