8. Troubleshooting
58
No.
Error message
Possible cause
Action
11 Periodic check
overdue.
The alarm appears when
the date for check or
calibration is overdue.
Periodic check schedule is
overdue, so check or
calibrate now, and you
can cancel the alarm. The
message will disappear.
When day interval is
changed,
it
will
be
effective only after check
or calibration is finished.
12 Periodic check
overdue.
Do you wish to
continue?
The message appears if
you attempt to continue
without
check
or
calibration.
Periodic check schedule is
overdue, so check or
calibrate now, and you
can cancel the alarm. The
message will disappear.
When day interval is
changed,
it
will
be
effective only after check
or calibration is finished.
13 X days left until
periodic check.
The message appears
when the due date is
approaching.
Check or calibrate now,
and you can cancel the
alarm. The message will
disappear.
When
day
interval is changed, it will
be effective only after
check or calibration is
finished.
14 Periodic Check
-- Today!
The alarm appears when
it is the day of periodic
check or calibration.
Check or calibrate now,
and you can cancel the
alarm. The message will
disappear.
When
day
interval is changed, it will
be effective only after
check or calibration is
finished.
15 Nozzle error.
Clear clog.
The alarm appears when
the nozzle of sample
changer
does
not
maneuver.
Remove any obstacle. If
the nozzle is bent or
damaged, change it.
Press [Stop] to cancel the
error.
16 Table error.
Set table.
The alarm appears when
the turntable does not
revolve
in
normal
condition.
Remove any obstacle.
Press [Stop] to cancel the
error.