8. Troubleshooting
60
Sample type
Technical tips
Cell window mists.
When running for many
hours (e.g. 24h) while
ambient temperature and
humidity are high (e.g.
35
°
C, 85%RH) and the cell
temperature is low (e.g.
15
°
C), the cell window may
mist.
Blow dry air through the
air intake vent for about 10
minutes.
Hint
If the problem with density measurement persists
even after the above countermeasure taken, contact
your local distributor.
Dry air
intake vent