12
3
Preparation
Fill the puncture unit with water
1. Use the positioning pillow to set the model to semi-sitting position for the training.
2. Connect the syringe filled with approximately 50 mL of water to the water supply tube. Open
the cock of the three-way tap on the water drainage tube and put the end of the tube into the
plastic cup.
2. Open the tube cock on the water supply tube and inject the water of the syringe into the
puncture unit. Inject water until water comes from the water drainage tube. Then close the
tube cock on the water supply tube and remove the syringe. And then set the cock of the
three-way tap on the water drainage tube into the position that is indicated in the picture below,
and then remove the tube from the plastic cup. Now the simulator is ready for training.
Open the cock
Open the tube cock
In case no water flow out after injecting 50mL of water, refill the syringe and inject
additional water into the puncture unit. Always be sure to close the cock on the
water supply tube before removing the syringe. Otherwise, water in the unit will be
drained due to atmospheric pressure.
Caution
Fill the Puncture Unit with Water [2]