8
3
Preparation
Do not remove the check valve on the three-way tap. If it is removed, it becomes
impossible to aspire water while training.
Always set the cock of the three-way tap into the position that is indicated in the
upper right picture. If the cock is not in this position, it becomes impossible to
aspire water while training.
When no water comes out from the water
drainage tube even if the water is supplied
from the irrigation bottle, lift the water
supply tube so that water left in the tube
flow into the pad.
When this does not work, pour a small
amount of water into the irrigation bottle
and open the cock on the tube again.
Then conduct the procedure in step (3)
after seeing water flowing out of the
drainage tube.
Caution
Caution
(2) After pouring water in the irrigation bottle, open the cock of the three-way tap on the water
drainage tube, and put the end of the water tube into the plastic cup.
(3) Open the cock on the water supply tube. The water in the irrigation bottle flows in the
puncture pad. The amount of water to fill the puncture unit is 500 mL. When water comes
out from the water drainage tube, close the cock on the water supply tube.
Set the cock of the three-way tap on the water drainage tube into the position that is
indicated in the picture below, and then remove the tube from the plastic cup. Now the
simulator is ready for training.
Open the cock
Open the cock
countercurrent
prevention valve
(check valve)
Fill the Puncture Unit with Water
Fill the puncture unit with water
[1]
Summary of Contents for MW15