Kusam-meco KM 175D Operation Manual Download Page 12

®

®

08

15

200.0 

m

F

2500 

m

F

±(2.0%rdg + 4dgts)

CAPACITANCE

1)

Accuracy

Resolution

Range

0.1 

m

F

1    

m

F

1)

 Accuracies with film capacitor or better

-40.0ºC ~ 99.9ºC

TEMPERATURE

Range

100.0ºC ~ 400ºC

-40.0ºF ~ 99.9ºF

100ºF ~ 752ºF

1.0% ~ 0.8ºC

Accuracy

1) 2)

1.0% ~ 1ºC

1.0% ~ 1.5ºF

1.0% ~ 2ºF

1) 

2) 

K-type thermocouple range & accuracy not included
Accuracies assume meter interior has the same temperature of the ambient

  (isothermal stage) for a correct junction voltage compensation.  Allow enough
time to reach the isothermal stage for a signficant change of ambient 
temperature.  It can take up to an hour for changes > 5°C.

HZ LINE LEVEL FREQUENCY

Accuracy :

 

1)

 DC-bias, if any, not more than 50% of Sine RMS.

1% + 5d

Function

600   V

1)

Sensitivity (Sine RMS)

50 V

5.00Hz~999.9Hz

Range

80ms PEAK-RMS for Clamp-on ACA 

Response

80ms to > 90% of specifications, 

& is specified from 2% of range.

CAUTION :

Using  Resistance,  Continuity  function  in  a  live  circuit  will
produce  false  results  and  may  damage  the  meter.  In  many  cases 
the  suspected  component(s) must be disconnected from the circuit
to obtain an accurate measurement reading.

W

 Resistance &     Continuity  functions :

Inputs    are    made    through    the    test    lead    terminals   

COM/+

Defaults  at  

W

 Resistance

.  Press  

SELECT 

 button momentarily to

toggle between the subject functions.

    Continuity function is convenient for checking wiring connections
&  operation  of  switches.    The  meter  provides Audible  as  well  as 
visible results.  A continuous beep tone, together with two flashing
LCD icons    & 

W

, indicates a complete wire.  This further improves

ease of use especially in noisy working environments.

18.68

 

MAX 600A

600V CAT III
300V CAT IV

V

W

Hz
EF

V

OFF

302.8

P-RMS

SELECT

HOLD

COM

+

CAT III

600V

CAT IV

300V

m

A

A

/

Duo

W

AUTO

AUTO

k

W

000. 1

 

MAX 600A

600V CAT III
300V CAT IV

V

W

Hz
EF

V

OFF

23.05

P-RMS

SELECT

HOLD

COM

+

CAT III

600V

CAT IV

300V

m

A

A

/

Duo

AUTO

W

1

SELECT

Summary of Contents for KM 175D

Page 1: ...rature 20 C 60 C 80 R H with battery removed o o o o o Temperature Coefficient Nominal 0 15 x specified accuracy C 0 C 18 C or 28 C 40 C or otherwise specified Power Supply Standard 1 5V AAA Battery x...

Page 2: ...52 F 1 0 0 8 C Accuracy1 2 1 0 1 C 1 0 1 5 F 1 0 2 F 1 2 K type thermocouple range accuracy not included Accuracies assume meter interior has the same temperature of the ambient isothermal stage for a...

Page 3: ...ometer Distance Meter Digital Lux Meter Network Cable Tester Power Factor Regulator Earth Resistance Tester Digital Panel Meters DC Power Supplies High Voltage Detector Calibrators Gas Analysers Frequ...

Page 4: ...dent or abnormal conditions of operation or handling KUSAM MECO authorized dealer shall extend this warranty on new and unused products to end user customers only but have no authority to extend a gre...

Page 5: ...WARNING To reduce the risk of fire or electric shock do not expose this product to rain or moisture The meter is intended only for indoor use Keep your hands fingers behind the hand finger barriers o...

Page 6: ...ve conductors But still individual protective equipment must be used if hazardous live parts in the installation where measurement is to be carried out could be accessible CAUTION Disconnect the test...

Page 7: ...range Backlighted LCD display Duo Core to simultaneously measure clamp on Amps A DMM functions VFD V Hz for fundamental V Hz of most Variable Frequency Drives Display Hold Non Contact EF Detection NC...

Page 8: ...abled automatically in this mode Hold feature freezes the display for later view LCD turns on Press the HOLD button momentarily to toggle the hold feature The Complementary Beeper feature can be selec...

Page 9: ...lead terminals COM Application Notes on Flame Sensors The DCmA function is designed especially for HVAC R flame sensor applications The 0 1 mA resolution is useful for identifying the minute current...

Page 10: ...nt 0 60A kV 50 60Hz 3 Specified accuracy is for measurements made at the jaw center When the conductor is not positioned at the jaw center add 2 to specified accuracy for position errors 4 Induced non...

Page 11: ...must be disconnected from the circuit to obtain an accurate measurement reading Capacitance Diode functions Inputs are made through the test lead terminals COM Defaults at Capacitance Press SELECT bu...

Page 12: ...from 2 of range CAUTION Using Resistance Continuity function in a live circuit will produce false results and may damage the meter In many cases the suspected component s must be disconnected from the...

Page 13: ...0 W W Response Time 32ms approx 3 000 V 1 5 rdg 5dgts DIODE TESTER Accuracy Resolution Range 1 mV Test Current 0 3mA typically Open Circuit Voltage 3 5VDC typical Non Contact EF Detection NCV An anten...

Page 14: ...00V CAT III 300V CAT IV V W Hz EF V OFF AUTO H P RMS 8 8 8 8 mFA V VFD P RMS SELECT HOLD COM CAT III 600V CAT IV 300V 2 3 1 4 6 10 mA A AUTO kW Hz 6 1 7 5 Duo The detected Electric Field strength is i...

Page 15: ...much as possible to get the best measuring accuracy For removal press the jaw trigger and remove the jaws from the conductor s CAUTION Adjacent current carrying devices such as transformers motors an...

Reviews: